New method to remove a broken guide pin in the hip joint.

نویسندگان

  • Mohammad Taghi Peivandi
  • Amir Reza Kachooei
  • Zohreh Nazemian
چکیده

Removing a broken guide wire is difficult and challenging. This article reports a method that was used to successfully remove a broken guide wire in 3 patients who underwent operative treatment for intertrochanteric fractures of the femur. Under general anesthesia, the first patient was placed on the orthopedic fracture table, and closed reduction of the hip fracture was performed under the image intensifier. A 2-mm non-threaded pin was used in the center of the femoral neck, and the amount of reaming required was determined. The reaming proceeded without any difficulty until the reamer failed to progress, which was attributed to the presence of a sclerotic or calcar region. Under fluoroscopy, on an anteroposterior view, the pin was parallel with the reamer; however, the authors failed to check the lateral image. After reaming through the hard tissue, the reamer suddenly passed through easily. On checking under the scope, it was noted that the pin was broken. The tip of the pin passed the acetabulum but did not enter the pelvis. Efforts to remove the broken pin failed. An attempt to reach the pin using a dynamic hip screw reamer also failed. Ultimately, the broken pin was removed using a 2-mm cannulated drill bit under fluoroscopy. After the experience with the first patient, in the other 2 cases, the authors used a cannulated drill bit to avoid any extensive reaming and trauma to the tissues. With this method, the authors were able to remove the broken pin easily in a few minutes.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture

Breakage of DHS guide wire during surgery and its migration into the pelvis through the hip joint is a rare complication and its removal can be very challenging for the surgeon. We share our experience of a similar case wherein we used an ‘iliofemoral’ approach to successfully remove the broken transfixing guidewire from the hip joint. Although iliofemoral approach is similar to the lateral win...

متن کامل

Intrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture

  During fixation of a femoral neck fracture in a 23-year-old male patient with cannulated screws, a broken guide wire fragment inadvertently advanced through the hip joint and protruded into the pelvis. A second surgical approach was needed to remove the broken fragment from the pelvis. Awareness of such a potentially devastating complication will make surgeons more cautious during implementa...

متن کامل

Migration of a Broken Steinman Pin into the Posterior Compartment of the Leg 12 Years after Total Hip Arthroplasty

This report is on a migration of a Steinman pin into the posterior com Department of the calf. It was used to fix a greater trochanteric osteotomy in a total hip arthroplasty. The puzzling note is how this metal piece passed through the anterolateral compartment of the thigh to the posterior compartment of the calf. Accord ing to our literature review, migration of fixation pins through the kne...

متن کامل

A technique to remove a broken guide wire transfixing the hip joint.

A broken guide wire while operating on hip is one of every surgeon's most dreaded nightmares. We report here a method which was used successfully to retrieve a broken guide wire from a patient who was being operated for core decompression and fibular grafting for avascular necrosis of the femoral head.

متن کامل

Intrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture.

During fixation of a femoral neck fracture in a 23-year-old male patient with cannulated screws, a broken guide wire fragment inadvertently advanced through the hip joint and protruded into the pelvis. A second surgical approach was needed to remove the broken fragment from the pelvis. Awareness of such a potentially devastating complication will make surgeons more cautious during implementatio...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Orthopedics

دوره 34 10  شماره 

صفحات  -

تاریخ انتشار 2011