18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register

نویسندگان

  • Birgitte Espehaug
  • Ove Furnes
  • Lars B Engesæter
  • Leif I Havelin
چکیده

BACKGROUND AND PURPOSE Few studies have compared the long-term survival of cemented primary total hip arthroplasties (THAs), and several prostheses have been used without adequate knowledge of their endurance. We studied long-term outcome based on data in the Norwegian Arthroplasty Register. PATIENTS AND METHODS The 10 most used prosthesis brands in 62,305 primary Palacos or Simplex cemented THAs reported to the Register from 1987 through 2007 were included. Survival analyses with revision as endpoint (for any cause or for aseptic loosening) were performed using Kaplan-Meier and multiple Cox regression with time-dependent covariates. Revision rate ratios (RRs) were estimated for the follow-up intervals: 0-5, 6-10, and > 10 years. RESULTS 5 prosthesis brands (cup/stem combinations) (Charnley, Exeter, Titan, Spectron/ITH, Link IP/Lubinus SP; n = 24,728) were investigated with 0-20 year follow-up (inserted 1987-1997). After 18 years, 11% (95% CI: 10.6-12.1) were revised for any cause and 8.4% (7.7-9.1) for aseptic loosening. Beyond 10 years of follow-up, the Charnley cup had a lower revision rate due to aseptic loosening than Exeter (RR = 1.8) and Spectron (RR = 2.4) cups. For stems, beyond 10 years we did not find statistically significant differences comparing Charnley with Titan, ITH, and SP stems, but the Exeter stem had better results (RR = 05). 10 prosthesis brands (9 cups in combination with 6 stems; n = 37,577) were investigated with 0-10 years of follow-up (inserted from 1998 through 2007). The Charnley cup had a lower revision rate due to aseptic loosening than all cups except the IP. Beyond 5 years follow-up, the Reflection All-Poly cup had a 14 times higher revision rate. For stems, beyond 5 years the Spectron-EF (RR = 6.1) and Titan (RR = 5.5) stems had higher revision rates due to aseptic loosening than Charnley. The analyses also showed a marked improvement in Charnley results between the periods 1987-1997 and 1998-2007. INTERPRETATION We observed clinically important differences between cemented prosthesis brands and identified inferior results for previously largely undocumented prostheses, including the commonly used prosthesis combination Reflection All-Poly/ Spectron-EF. The results were, however, satisfactory according to international standards.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Improved results of primary total hip replacement

BACKGROUND AND PURPOSE Over the past 20 years, several changes in treatment policy and treatment options have taken place regarding hip replacement. For this reason, we wanted to investigate the results after hip replacement in terms of revision rate, during a 21-year period among hip replacements reported to the Norwegian Arthroplasty Register. METHODS 110,882 primary total hip replacements ...

متن کامل

Failure rates for 4762 revision total hip arthroplasties in the Norwegian Arthroplasty Register.

We present the results for 4762 revision total hip arthroplasties with no previous infection in the hip, which were reported to the Norwegian Arthroplasty Register between 1987 and 2003. The ten-year failure rate for revised prostheses was 26% (95% CI 25 to 26). Cox regression analyses were undertaken separately for acetabular and femoral revision components. Cemented revision components withou...

متن کامل

257 ankle arthroplasties performed in Norway between 1994 and 2005.

BACKGROUND AND PURPOSE There have been few reports on the long-term outcome of ankle replacements. The Norwegian Arthroplasty Register has been registering ankle replacements since 1994, but no analysis of these data has been published to date. Here we report data on the use of total ankle replacements and the revision rate in the Norwegian population over a 12-year period. METHODS We used th...

متن کامل

Early aseptic loosening of uncemented femoral components in primary total hip replacement. A review based on the Norwegian Arthroplasty Register.

The Norwegian Arthroplasty Register recorded 24,408 primary total hip replacements from 1987 to 1993; 2907 of them (13%) were performed with uncemented femoral components. We have compared the results of eight different designs, each used in more than 100 patients. Survivorship of the components was estimated by the Kaplan-Meier method using revision for aseptic loosening of the femoral compone...

متن کامل

Hemiarthroplasties after hip fractures in Norway and Sweden: a collaboration between the Norwegian and Swedish national registries.

National registration of hemiarthroplasties after hip fractures has been established in both Norway and Sweden. We aimed to investigate differences in demographics, choice of implant selection, surgical approaches, and reoperations between the Norwegian Hip Fracture Register (NHFR) and the Swedish Hip Arthroplasty Register (SHAR). As part of the Nordic Arthroplasty Register Association (NARA) p...

متن کامل