Measurement of femoral head penetration in polyethylene using a 3-dimensional CT-scan technique
نویسندگان
چکیده
BACKGROUND Current techniques for measuring in vivo polyethylene wear suffer from a range of problems, resulting in an unacceptable lack of repeatability and/or insufficient accuracy when they are used to measure the low wear rates associated with new, highly crosslinked polyethylene. We describe an improved CT method for measurement of 3D femoral head penetration in PE acetabular cups that has sufficient accuracy and repeatability to allow assessment of the wear potential of modern implants. METHOD The accuracy and repeatability of the CT-scan method was determined by blindly repeating measurements on a precisely calibrated 28-mm prosthetic head and by comparing them with direct metrological measurements on 10 acetabular specimens with in vitro wear from machining, and on 8 explanted acetabular specimens with in vivo wear. RESULTS The intra- and interobserver errors in femoral head diameter were 0.036 mm (SD 0.044) and 0.050 mm (SD 0.022), respectively. CT estimated femoral head penetration in both all-poly and metal-backed acetabular components with accuracy ranging from 0.009 to 0.245 mm (mean 0.080; SD 0.067). INTERPRETATION We found that the CT method is rapid, is accurate, and has repeatability and ease of availability. Using a slice thickness of 0.0625 mm, this method can detect wear—and also the threshold for the wear rate that causes osteolysis—much earlier than previous methods.
منابع مشابه
Validation of a 3D CT method for measurement of linear wear of acetabular cups
BACKGROUND We evaluated the accuracy and repeatability of a 3D method for polyethylene acetabular cup wear measurements using computed tomography (CT). We propose that the method be used for clinical in vivo assessment of wear in acetabular cups. MATERIAL AND METHODS Ultra-high molecular weight polyethylene cups with a titanium mesh molded on the outside were subjected to wear using a hip sim...
متن کاملComparison of in vivo wear between polyethylene liners articulating with ceramic and cobalt-chrome femoral heads.
At yearly intervals we compared the radiological wear characteristics of 81 alumina ceramic femoral heads with a well-matched group of 43 cobalt-chrome femoral heads. Using a computer-assisted measurement system we assessed two-dimensional penetration of the head into the polyethylene liner. We used linear regression analysis of temporal data of the penetration of the head to calculate the true...
متن کاملAn Improved Technique Using the Martell Method with Ellipse Fitting: a Comparitive Study Using a Minimum 5 Year Follow-up of Highly Cross-linked Polyethylene
+*Bragdon, C.R; *Greene, M.E; **Martell, J.M; *Geller, J.A.; Kwong, M; *Harris, W.H.; *Malchau, H. +*Orthopaedic Biomechanics and Biomaterials Laboratory, Massachusetts General Hospital, Boston, MA 02114 617-726-3869 fax: 617-726-3883 [email protected] Introduction Highly cross-linked polyethylene has been used as the bearing surface of total hip arthroplasty for over 6 years. Several repor...
متن کاملValidation of a new 2-D technique for radiographic wear measurement of cemented, highly cross-linked polyethylene acetabular cups
A new 2-D radiographic wear measurement system has been developed which enables the low wear of highly cross-linked polyethylene acetabular cups to be accurately and precisely measured from standard, pelvis radiographs. The software was validated using radiographic images of a measurement jig which could vary the cup orientation and simulate the effect of pelvic tilt/rotation. Wear was simulate...
متن کاملWear of a sequentially annealed polyethylene acetabular liner
BACKGROUND AND PURPOSE We previously reported on a randomized controlled trial (RCT) that examined the effect of adding tobramycin to bone cement after femoral stem migration. The present study examined femoral head penetration into both conventional and highly crosslinked polyethylene acetabular liners in the same group of RCT patients, with a minimum of 5 years of postoperative follow-up. P...
متن کامل