Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner
نویسندگان
چکیده
PURPOSE To compare the precision of anterior chamber angle (ACA) and anterior chamber depth (ACD) measurements taken with ultrasound biomicroscopy (UBM) and the Artemis-2 Very High Frequency Ultrasound Scanner (VHFUS) in normal subjects. DESIGN Prospective study. METHODS We randomly selected one eye from each of 59 normal subjects in this study. Two subjects dropped out of the study; the associated data were excluded from analysis. ACA and ACD measurements were obtained using the VHFUS and the UBM. The results were compared statistically using repeated-measures analysis of variance for the intraobserver repeatability, unpaired t-test, and limits of agreement. RESULTS The average ACA values for the UBM and the VHFUS (±standard deviation) were 41.83° ± 5.03° and 33.36° ± 6.03°, respectively. The average ACD values were 2.96 ± 0.34 mm and 2.87 ± 0.31 mm. The intraobserver repeatability analysis of variance P-values for ACA and ACD measurements using UBM were 0.10 and 0.68, respectively; for the Artemis-2 VHFUS, the respective values were 0.68 and 0.09. The difference in ACA measurements was statistically significant (t = 8.41; P < 0.0001), while the difference in ACD values was not (t = 1.51; P < 0.13). The mean ACA difference was 8.50° ± 2.50°, and the limits of agreement were +13.30° to -3.60°. The mean ACD difference was 0.09 ± 0.27 mm, and the limits of agreement ranged from 0.61 mm to -0.43 mm. The mean difference percentage of ACD was 3.1% for both instruments. CONCLUSION In case of the ACD, both instruments can be used interchangeably; however, with the ACA instruments, they cannot be used interchangeably.
منابع مشابه
Measurements of central corneal thickness using two immersion ultrasound techniques and optical technique.
OBJECTIVE To compare the accuracy of central corneal thickness measurements using ultrasound biomicroscopy, Orbscan II tomography and an Artemis-2 very high frequency ultrasound scanner. METHODS The prospective study was conducted at Eye World Centre, Riyadh, Saudi Arabia, from September to November 2012. One eye from each of 60 normal subjects was analysed. The central corneal thickness was ...
متن کاملAgreement between Orbscan II, VuMAX UBM and Artemis-2 very-high frequency ultrasound scanner for measurement of anterior chamber depth
BACKGROUND The aim was to compare the anterior chamber depth (ACD) measurements taken with Orbscan II, ultrasound biomicroscopy (UBM) and the Artemis-2 VHF (very-high-frequency) ultrasound scanner in normal subjects. METHODS In this prospective study, one eye from each of 60 normal subjects was randomly selected. Three subjects dropped out of the study because they were apprehensive about the...
متن کاملComparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes
PURPOSE To compare the precision of central corneal thickness (CCT) measurements taken with the handheld ultrasound pachymeter (USP), ultrasound biomicroscopy (UBM), and the Artemis-2 very high frequency ultrasound scanner (VHFUS) on normal subjects. DESIGN Prospective study. METHODS One eye from each of 61 normal subjects was randomly selected for this study. The measurements of the CCT we...
متن کاملUltrasound Biomicroscopy in Glaucoma: An Update
UBM is a high-frequency ultrasound technology that provides exceptionally detailed two-dimensional gray-scale images of the various anterior segment structures and evaluates them both quantitatively and qualitatively. It allows noninvasive in vivo imaging of structural details of the anterior ocular segment at near microscopic resolution and thus, there are many other applications of this new i...
متن کاملEvaluation of Anterior Segment Changes Following Laser Peripheral Iridotomy Using Pentacam Scheimpflug Imaging System in Eyes with Primary Angle Closure (PAC)
Primary angle-closure glaucoma (PACG) is a major form of glaucoma in Asia. It is a leading cause of blindness worldwide. Laser peripheral iridotomy (LPI) is the standard firstline intervention in both acute and chronic forms. It is successful in preventing recurrence of acute attacks and virtually eliminates the risk of an acute attack in the fellow eye. Peripheral iridotomy eliminates pupillar...