Evaluation of Hylamer-zirconia hip arthroplasty in patients less than 50 years of age.

نویسندگان

  • John G Kennedy
  • Padhraig F O'Loughlin
  • James A Harty
  • Kevin Casey
  • David Zurakowski
  • William B Quinlan
چکیده

The purpose of this prospective study was to determine the clinical outcomes of patients who had total hip arthroplasties with a zirconia (Zr) - Hylamer pairing. Fifty-eight consecutive cemented total hip arthroplasties with a Hylamer cup and 22.25 mm Zr head were examined for component wear and failure. The SF-36 and Mayo hip score were used to evaluate patient outcome. Regression and Kaplan Meier survival analyses were used to determine outcome. The mean Mayo postoperative score was 94 points. The median post-operative SF-36 score was 84 points. The mean rate of wear was 0.097 +/- 0.02 mm/year, with no evidence of loosening or significant osteolysis at a minimum follow-up of 5 years. Kaplan-Meier survival analysis was 98% at 3-year follow-up and 94% at 5-year follow-up. This study has confounded previous reports relating to Zirconia-Hylamer counter-surface components.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Wear of enhanced ultra-high molecular-weight polyethylene (Hylamer) in combination with a 22.225 mm diameter zirconia femoral head.

We have prospectively studied the wear of enhanced ultra-high molecular-weight polyethylene (Hylamer) in combination with a zirconia femoral head of 22.225 mm diameter on a cemented, triple-tapered, collarless, polished stem, the C Stem. The 71 patients who underwent total hip arthroplasty had a mean follow-up of six years (3 to 8). No patient died or was lost to follow-up. The clinical results...

متن کامل

Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head.

We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia ceramic head and an acetabular component of cross-linked ultra-high molecular-weight polyethylene (Hylamer). Between 1995 and 1999 we implanted 29 hips in 26 patients with a mean age of 49.2 years. Survivorship analysis in this group revealed a failure rate of 67.6% at...

متن کامل

A comparison of polyethylene wear in hips with cobalt-chrome or zirconia heads.

We have reviewed 70 patients with bilateral simultaneous total hip arthroplasties to determine the rate of failure and to compare polyethylene wear and osteolysis between an implant with a cobalt-chrome head and Hylamer liner with that of a zirconia head and Hylamer liner. The mean thickness of the polyethylene liner was 11.0 mm (8.8 to 12.2) in the hip with a zirconia head and 10.7 mm (8.8 to ...

متن کامل

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner.

Clinical results after total hip arthroplasty with the use of a Hylamer liner have been described as being poor. In this study, polyethylene wear rates as well as clinical results were assessed after hip arthroplasty using a Hylamer liner. Thirty patients were followed radiographically and clinically after total hip arthroplasty with a Hylamer liner. The mean follow-up was six years. Radiolucen...

متن کامل

In-Hospital Outcomes after Hemiarthroplasty versus Total Hip Arthroplasty for Isolated Femoral Neck Fractures

  Background:    Previous studies suggest total hip arthroplasty may have some benefits compared to hemi-arthroplasty for displaced intracapsular femoral neck fractures in patients more than 60 years of age. The primary research question of our study was whether in-hospital adverse events, post-operative length of stay (LOS) and mortality in patients 60 year of age or older differed between tot...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Acta orthopaedica Belgica

دوره 74 4  شماره 

صفحات  -

تاریخ انتشار 2008