Aneurysm sac pressure monitoring: effect of technique on interpretation of measurements.

نویسندگان

  • J W Hinnen
  • M J T Visser
  • J H van Bockel
چکیده

OBJECTIVES The purpose of this study is to determine the accuracy of measuring pressure with a fluid filled pressure device (needle) and a non-fluid filled pressure device (catheter) inside a thrombosed aneurysmal sac after exclusion from circulation by endovascular grafting. METHODS In a static environment, consisting of a syringe to which a pressure monitoring kit was connected, experiments were performed to study the influence of the type of device (either needle or catheter) and the effect of the characteristics of the medium on the accuracy and reproducibility of pressure measurements. The pressures obtained with the needle in the different kinds of media were compared with those obtained in blood. Similar experiments were performed using a pressure catheter. Subsequently, pressure measurements were performed in a dynamic and physiological environment. This environment consisted of an artificial circulation in which an aneurysm, constructed of porcine aorta and filled with human aortic thrombus, was mounted. The pressures were compared and analyzed by Bland-Altman plots. RESULTS Under static conditions, the pressure levels obtained by a needle in blood, starch solution and thrombus were similar. Under identical conditions, pressures obtained by a catheter in starch solution were significant lower than the pressures measured in blood (P<0.05). Under dynamic pressure conditions the reproducibility of pressures obtained with the needle inserted in the human thrombus was very poor. CONCLUSION A needle pressure measuring device, inserted into media like human fibrin thrombus, does not yield accurate and reproducible results. A catheter with a tip-sensor rather than a needle is superior to study the pressure in the aneurysm sac after EVAR.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of intraluminal thrombus on pressure transmission in the abdominal aortic aneurysm.

OBJECTIVE To clarify the effect of intraluminal thrombus on pressure transmission. METHODS A saccular aneurysm was inserted into an artificial circulation system. Subsequently, the saccular aneurysm was filled with eight different human aortic aneurysm thrombus samples. Starch solution in an empty aneurysm was used as a control. A pressure sensor measured the pressure in the circulation, and ...

متن کامل

Re: "Aneurysm sac pressure monitoring: effect of pulsatile motion of the pressure sensor on the interpretation of measurements".

PURPOSE To determine the effect of pulsatile motion of graft-attached pressure sensors on the accuracy of aneurysm sac pressure measurement. METHODS Pressure inside a pressure box was measured with a sensor attached to a stent-graft (Sensorgraft) facing a sensor in fixed position (Sensorbox). Maximum inter-sensor variation of Sensorgraft and Sensorbox was determined in static experiments. Exp...

متن کامل

Correlation Between Intrasac Pressure Measurements of a Pressure Sensor and an Angiographic Catheter During Endovascular Repair of Abdominal Aortic Aneurysm

PURPOSE To establish a correlation between intrasac pressure measurements of a pressure sensor and an angiographic catheter placed in the same aneurysm sac before and after its exclusion by an endoprosthesis. METHODS Patients who underwent endovascular abdominal aortic aneurysm repair and received an EndoSure wireless pressure sensor implant between March 19 and December 11, 2004 were enrolle...

متن کامل

Pressure changes within the sac of human cerebral aneurysms in response to artificially induced transient increases in systemic blood pressure.

Formation and rupture of cerebral aneurysms have been associated with chronic hypertension. The effect of transient increase in blood pressure and its effect on intra-aneurysmal hemodynamics have not been studied. We examined the effects of controlled increases in blood pressure on different pressure parameters inside the sac of human cerebral aneurysms and corresponding parent arteries using i...

متن کامل

Durability and validity of a remote, miniaturized pressure sensor in an animal model of abdominal aortic aneurysm.

PURPOSE To investigate whether a remote, miniaturized pressure sensor could maintain calibration and function through organized thrombus over an extended period in a porcine model of abdominal aortic aneurysm (AAA). METHODS Six adult pigs had an AAA surgically created and excluded. A sensor zeroed to atmospheric pressure was placed within the aneurysm sac and another within the suprarenal aor...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery

دوره 29 3  شماره 

صفحات  -

تاریخ انتشار 2005