X-ray imaging detector based on a position sensitive THCOBRA with resistive line
نویسندگان
چکیده
A new X-ray imaging detector based on a 2D-THCOBRA micropatterned structure using a simple position readout is proposed. It consists of a hybrid device that combines the properties of a THGEM and a 2D-MHSP in a single structure, having two charge multiplication stages reaching the demanded gains for the use of charge division readout methods. For position determination, the new 10×10 cm2 2D-THCOBRA structure uses two orthogonal resistive lines located at the end of the anode and top electrodes. The charge signal pulses collected at the end of each resistive line are digitized and processed in order to determine the center-ofgravity of the electron avalanche distribution. The detector uses a preamplification stage performed by a THGEM, followed by the 2DTHCOBRA, being operated in Ne/5%CH4 at atmospheric pressure. A characterization of the detector in terms of charge gain was made as a function of the voltage applied to the electrodes of the 2D-THCOBRA structure. The energy resolution of the system was also measured, envisaging Energy Weighting Technique (EWT) applications. The system characterization in terms of spatial resolution is presented together with imaging examples to evaluate its performance in X-ray imaging applications.
منابع مشابه
طراحی، شبیهسازی و ساخت یک نمونه اتاقک گازی پرتو ایکس حساس به موقعیت
In this paper a one dimentional position-sensitive X-ray gas detector has been designed, simulated and constructed based on Multi Wire Proportional Chamber (MWPC) with delay line position readout. These kinds of detectors are useful in soft X-ray imaging and are capable of being extended to two dimensions position readout easily. The position resolution of this detector is estimated to be 230µm...
متن کاملA Method for Reduction of Parallax Broadening in Gas-Based Position Sensitive Detectors
A description is given of the principle and experimental verification of a new method which significantly reduces broadening of the position response, due to parallax, for radiation incident on a detector at finite angles of incidence. The technique, which can be implemented in gas-based position-sensitive detectors with planar geometry, substantially improves position resolution for scattering...
متن کاملDesign Consideration on Position Sensitive Detectors Based on LuAG:Pr Scintillators for High Energy X-ray Cargo Inspection
To prevent illegal trade of weapons, X-ray scanning systems are used for non-intrusive inspection of import containers and cargoes at every seaport and airport. Conventional X-ray scanning system for cargoes is based mainly on X-rays of several MeV energy generated by an electron Linac. For high quality X-ray imaging, an X-ray source with good directionality, narrow width of energy spectrum and...
متن کاملCvd-diamond-based Position Sensitive Photoconductive Detector for High-flux X-rays and Gamma Rays
A position-sensitive photoconductive detector (PSPCD) using insulating-type CVD diamond as its substrate material has been developed at the Advanced Photon Source (APS). Several different configurations, including a quadrant pattern for a x-ray-transmitting beam position monitor (TBPM) and 1-D and 2-D arrays for PSPCD beam profilers, have been developed. Tests on different PSPCD devices with hi...
متن کاملA Position Sensitive Detector for 1 - 10 MeV Gamma RaysJoel
A position sensitive detector for 9 MeV gamma rays has been developed as part of a contraband detection system for airports. The system images contraband, such as explosives or drugs, in a geometry similar to medical x-ray computed tomography. A point source of gamma rays shines through an unknown object. Transmission projections of the areal density of detected gamma rays are reconstructed int...
متن کامل