Evaluation of evacuated blood-collection tubes: effects of three types of polymeric separators on therapeutic drug-monitoring specimens.
نویسندگان
چکیده
The potential of three types of separator materials found in conventional blood-collection tubes for interference in therapeutic drug measurements was assessed. None of the separators (based on acrylic, silicone, or polyester polymers) had any significant effect on the concentrations of seven drugs (theophylline, digoxin, phenytoin, phenobarbital, gentamicin, ethanol, and cyclosporine) in blood specimens that were processed and analyzed promptly. Storage of specimens for 24 h resulted in an average 2.4% increase in theophylline values in specimens collected in tubes with the acrylic separator (P = 0.024); an average 8.1% decrease in phenytoin in specimens collected in tubes with the polyester-based separator (P < 0.001); and an average 4.2% decrease in phenobarbital in specimens collected in tubes with the polyester-based separator (P = 0.02). All other drug concentrations were not significantly affected. A small decrease in phenytoin (7.9%; P < 0.01) was seen when the specimen volume in 7-mL tubes containing polyester-based separator was reduced to 1.0 mL; all other drug concentrations were unaffected by partial filling of tubes. Paired blood specimens from pediatric patients, when collected in plain tubes and tubes containing acrylic separator, yielded no significant differences for theophylline, digoxin, tobramycin, phenytoin, or phenobarbital concentrations. The three commercially available separators had only small effects on therapeutic drug concentrations, and a newly developed separator based on an acrylic resin was suitably inert.
منابع مشابه
Effects of contaminants in blood-collection devices on measurements of therapeutic drugs.
Substances in evacuated blood-collection devices produce gas-chromatographic peaks with retention times similar to those of drugs. All tubes tested except the serum separator tubes contained tris(2-butoxyethyl) phosphate, as determined by mass spectrometry. In addition, the partition coefficient during the solvent-extraction step increased for some drugs by as much as 40%, while for other drugs...
متن کاملCarbohydrate-deficient transferrin is not affected by serum separators.
We studied the possible effects on serum carbohydrate-deficient transferrin (CDT) determination by a CDTect (Pharmacia) method of serum isolation in four different types of blood-collection tubes, namely: (1) glass tubes (glass Vacutainer tubes with no additive); (2) S-Monovette Neutral tubes (plastic tubes with no additive); (3) S-Monovette Serum tubes (plastic tubes with kaolin-coated plastic...
متن کاملInterferences from blood collection tube components on clinical chemistry assays
Improper design or use of blood collection devices can adversely affect the accuracy of laboratory test results. Vascular access devices, such as catheters and needles, exert shear forces during blood flow, which creates a predisposition to cell lysis. Components from blood collection tubes, such as stoppers, lubricants, surfactants, and separator gels, can leach into specimens and/or adsorb an...
متن کاملUnsuitability of evacuated tubes for monitoring heparin therapy by activated partial thromboplastin time.
Activated partial thromboplastin times (APTT) for monitoring heparin therapy for venous thromboembolism tended to be inappropriately short if blood was collected in commercially available evacuated glass tubes. Five types of evacuated tubes marketed under the trade names Vacutainer and Venoject were examined. The APTT of heparinized blood collected in these tubes correlated poorly (r = 0.04 to ...
متن کاملTherapeutic Drug Monitoring of Sirolimus,Correlation With Laboratory Parameters In Transplant Patients
Sirolimus is a potent immunosuppressive agent administered as prophylactic agent to prevent rejection after organ transplantation. Sirolimus must be used within a narrow therapeutic window. Due to inter- and intra-variability, sirolimus blood concentrations may be affected, therefore, there is no possibility of predicting the sirolimus blood concentrations based on the dose patients received. T...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 39 8 شماره
صفحات -
تاریخ انتشار 1993