Propofol detection and quantification in human blood: the promise of feedback controlled, closed-loop anesthesia.

نویسندگان

  • Francine Kivlehan
  • Edward Chaum
  • Ernő Lindner
چکیده

The performance of a membrane-coated voltammetric sensor for propofol (2,6-diisopropylphenol) has been characterized in long term monitoring experiments using an automated flow analytical system (AFAS) and by analyzing human serum and whole blood samples by standard addition. It is shown that the signal of the membrane-coated electrochemical sensor for propofol is not influenced by the components of the pharmaceutical formulation of propofol (propofol injectable emulsion). The current values recorded with the electrochemical propofol sensor in buffer solutions and human serum samples spiked with propofol injectable emulsion showed excellent correlation with the peak heights recorded with an UV-Vis detector during the HPLC analysis of these samples (R(2) = 0.997 in PBS and R(2) = 0.975 in human serum). However, the determination of propofol using the electrochemical method is simpler, faster and has a better detection limit (0.08 ± 0.05 μM) than the HPLC method (0.4 ± 0.2 μM). As a first step towards feedback controlled closed-loop anesthesia, the membrane-coated electrochemical sensor has been implemented onto surface of an intravenous catheter. The response characteristics of the membrane-coated carbon fiber electrode on the catheter surface were very similar to those seen using a macroelectrode.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of closed-loop controlled administration of propofol using Bispectral Index as the controlled variable versus "standard practice" controlled administration.

BACKGROUND This report describes a new closed-loop control system for propofol that uses the Bispectral Index (BIS) as the controlled variable in a patient-individualized, adaptive, model-based control system, and compares this system with manually controlled administration of propofol using hemodynamic and somatic changes to guide anesthesia. METHODS Twenty female patients, American Society ...

متن کامل

Robust closed-loop control of induction and maintenance of propofol anesthesia in children.

BACKGROUND During closed-loop control, a drug infusion is continually adjusted according to a measure of clinical effect (e.g., an electroencephalographic depth of hypnosis (DoH) index). Inconsistency in population-derived pediatric pharmacokinetic/pharmacodynamic models and the large interpatient variability observed in children suggest a role for closed-loop control in optimizing the administ...

متن کامل

Migrating from target-controlled infusion to closed-loop control in general anaesthesia

The target-controlled infusion (TCI) technique has been successfully and commercially used in clinical general anaesthesia with the intravenous anaesthetic agent propofol. The technique is based on a population pharmacokinetic model and is an open-loop control system. Closed-loop control requires a reliable and consistent signal for feedback utilisation. With all anaesthetic agents the somatose...

متن کامل

Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System

BACKGROUND The CONCERT-CL closed-loop infusion system designed by VERYARK Technology Co., Ltd. (Guangxi, China) is an innovation using TCI combined with closed-loop controlled intravenous anesthesia under the guide of BIS. In this study we performed a randomized, controlled, multicenter study to compare closed-loop control and open-loop control of propofol by using the CONCERT-CL closed-loop in...

متن کامل

Closed-loop control of anesthesia using Bispectral index: performance assessment in patients undergoing major orthopedic surgery under combined general and regional anesthesia.

BACKGROUND The Bispectral Index (BIS) is an electroencephalogram-derived measure of anesthetic depth. A closed-loop anesthesia system was built using BIS as the control variable, a proportional-integral-differential control algorithm, and a propofol target-controlled infusion system as the control actuator. Closed-loop performance was assessed in 10 adult patients. METHODS Ten adult patients ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Analyst

دوره 140 1  شماره 

صفحات  -

تاریخ انتشار 2015