Portable pulse-dose oxygen concentrators should not be used with noninvasive ventilation.
نویسندگان
چکیده
An increasing number of patients, mainly COPD and overlap-syndrome patients, simultaneously use home oxygen therapy and noninvasive ventilation (NIV) at night. Usually the oxygen source is a stationary concentrator. We report a patient who, without a medical recommendation, was using a portable oxygen concentrator during nocturnal NIV. In the laboratory, with the patient, we tested the portable oxygen concentrator's triggering and oxygen delivery, with the supplemental oxygen connected at 3 different positions: near the ventilator, near the exhalation valve, and on the nasal mask port. We also tested the concentrator's triggering capacity by placing the nasal prongs and the nasal mask independently. We tested ventilator inspiratory pressures of 10, 14, and 18 cm H(2)O, and expiratory pressures of 4 and 6 cm H(2)O. The portable oxygen concentrator did not detect the patient's inspiratory effort or deliver the required oxygen flow at any of the tested settings. We recommended that the patient not use the portable oxygen concentrator during nocturnal NIV.
منابع مشابه
A comparative study of 3 portable oxygen concentrators during a 6-minute walk test in patients with chronic lung disease.
BACKGROUND The purpose of this study was to compare the ability of 3 portable oxygen concentrators (POCs) to maintain S(pO2) ≥ 90% during exercise in patients with chronic lung disease. METHODS Twenty-one subjects with chronic lung disease (18 with COPD, 3 with pulmonary fibrosis) and documented room air exertional S(pO2) ≤ 85% performed four 6-min walk tests: a control walk using the subject...
متن کاملEffect of the anatomic reservoir on low-flow oxygen delivery via nasal cannula: constant flow versus pulse flow with portable oxygen concentrator.
BACKGROUND The F(IO(2)) for a nasal cannula with constant flow (CF) depends on the anatomic reservoir (AR), which is affected by changes in frequency and end-expiratory flow. Conversely, pulse flow (PF) devices do not require the AR. The purpose of this study was to compare the F(IO(2)) delivered by a nasal cannula supplied by CF via oxygen tank with that delivered by PF delivered via portable ...
متن کاملEfficacy of a portable oxygen concentrator with pulsed delivery for treatment of hypoxemia during anesthesia of wildlife.
Portable battery-driven oxygen concentrators provide an alternative to the use of oxygen cylinders for treatment of hypoxemia during field anesthesia. The aim of this study was to evaluate the use of the EverGo Portable Oxygen Concentrator (Respironics, Murrysville, Pennsylvania 15668, USA) with pulse-dose delivery for improvement of arterial oxygenation during anesthesia of wildlife. This conc...
متن کاملOxygenation With a Single Portable Pulse-Dose Oxygen-Conserving Device and Combined Stationary and Portable Oxygen Delivery Devices in Subjects With COPD.
BACKGROUND Portable oxygen devices simplify and facilitate patient therapy. This study was designed to compare S(pO2) and patient satisfaction with a portable oxygen concentrator or a combined system consisting of a fixed device with continuous-flow oxygen dispensation and a portable device with pulse dispensation for ambulation. METHODS This crossover trial assessed 25 subjects with COPD (92...
متن کاملSupplemental oxygen needs during sleep. Who benefits?
The physiologic changes that occur in ventilation during sleep contribute to nocturnal oxygen desaturation in those with lung disease. Nocturnal supplemental oxygen is often used as therapy, although convincing data exist only for those who are hypoxemic both during sleep and wake. Ongoing trials may help address whether oxygen should be used in those with only desaturation during sleep. If use...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Respiratory care
 
دوره 56 12 شماره
صفحات -
تاریخ انتشار 2011