Pull-out strength of cemented solid versus fenestrated pedicle screws in osteoporotic vertebrae

نویسندگان

  • C. I. Leichtle
  • A. Lorenz
  • S. Rothstock
  • J. Happel
  • F. Walter
  • T. Shiozawa
  • U. G. Leichtle
چکیده

OBJECTIVES Cement augmentation of pedicle screws could be used to improve screw stability, especially in osteoporotic vertebrae. However, little is known concerning the influence of different screw types and amount of cement applied. Therefore, the aim of this biomechanical in vitro study was to evaluate the effect of cement augmentation on the screw pull-out force in osteoporotic vertebrae, comparing different pedicle screws (solid and fenestrated) and cement volumes (0 mL, 1 mL or 3 mL). MATERIALS AND METHODS A total of 54 osteoporotic human cadaver thoracic and lumbar vertebrae were instrumented with pedicle screws (uncemented, solid cemented or fenestrated cemented) and augmented with high-viscosity PMMA cement (0 mL, 1 mL or 3 mL). The insertion torque and bone mineral density were determined. Radiographs and CT scans were undertaken to evaluate cement distribution and cement leakage. Pull-out testing was performed with a material testing machine to measure failure load and stiffness. The paired t-test was used to compare the two screws within each vertebra. RESULTS Mean failure load was significantly greater for fenestrated cemented screws (+622 N; p ⩽ 0.001) and solid cemented screws (+460 N; p ⩽ 0.001) than for uncemented screws. There was no significant difference between the solid and fenestrated cemented screws (p = 0.5). In the lower thoracic vertebrae, 1 mL cement was enough to significantly increase failure load, while 3 mL led to further significant improvement in the upper thoracic, lower thoracic and lumbar regions. CONCLUSION Conventional, solid pedicle screws augmented with high-viscosity cement provided comparable screw stability in pull-out testing to that of sophisticated and more expensive fenestrated screws. In terms of cement volume, we recommend the use of at least 1 mL in the thoracic and 3 mL in the lumbar spine.Cite this article: C. I. Leichtle, A. Lorenz, S. Rothstock, J. Happel, F. Walter, T. Shiozawa, U. G. Leichtle. Pull-out strength of cemented solid versus fenestrated pedicle screws in osteoporotic vertebrae. Bone Joint Res 2016;5:419-426.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

May "Dubel" be a solution for pullout problem of the pedicle screws at osteoporotic spine?

AIM To improve the strength of stabilization systems currently used in osteoporotic spinal fractures, essentially by increasing the fixation force of pedicle screws. MATERIAL AND METHODS Six human cadaveric vertebrae were used. Bone mineral densities of the specimens were measured with Dual Energy X-ray Absorptiometry in order to assess the osteoporosis. All vertebrae were found to be severel...

متن کامل

Srovnávací analýza šroubů Socon CS a Socon v ošetření osteoporotických zlomenin torakolumbální páteře – biomechanická studie

PURPOSE OF THE STUDY To compare the anchorage strength of Socon CS cannulated pedicle screws (B. Braun, Aesculap) with that of Socon screws in human cadaver vertebrae, using pull-out strength testing. MATERIAL Twelve samples involving T12 to L3 vertebrae were prepared and a total of 20 pedicle screws, 10 Socon CS and 10 Socon screws, were inserted in them. All screws were 55 mm in length and 6 ...

متن کامل

Comparative Analysis of Effect of Density, Insertion Angle and Reinsertion on Pull-Out Strength of Single and Two Pedicle Screw Constructs Using Synthetic Bone Model

STUDY DESIGN Biomechanical study. PURPOSE To determine the effect of density, insertion angle and reinsertion on pull-out strength of pedicle screw in single and two screw-rod configurations. OVERVIEW OF LITERATURE Pedicle screw pull-out studies have involved single screw construct, whereas two screws and rod constructs are always used in spine fusions. Extrapolation of results using the si...

متن کامل

Axial pullout strength comparison of different screw designs: fenestrated screw, dual outer diameter screw and standard pedicle screw

BACKGROUND The pullout strength of pedicle screws is influenced by many factors, including diameter of the screws, implant design, and augmentation with bone cement such as PMMA. In the present study, the pullout strength of an innovative fenestrated screw augmented with PMMA was investigated and was compared to unaugmented fenestrated, standard and dual outer diameter screw. METHODS Twenty f...

متن کامل

The effect of lateral wall perforation on screw pull-out strength: a cadaveric study

BACKGROUND Lateral pedicle wall perforations occur frequently during pedicle screw insertion. Although it is known that such an occurrence decreases the screw pull-out strength, the effect has not been quantified biomechanically. MATERIALS AND METHODS Twenty fresh cadaveric lumbar vertebrae were harvested, and the bone mineral density (BMD) of each was evaluated with dual-energy radiography a...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 5  شماره 

صفحات  -

تاریخ انتشار 2016