A comparison of manifest refractions, cycloplegic refractions and retinoscopy on the RMA-3000 autorefractometer in children aged 3 to 15 years
نویسندگان
چکیده
PURPOSE The study was conducted to compare the accuracy of readings of the RMA-3000 autorefractometer (Topcon, Tokyo, Japan) with traditional retinoscopy as a means of determining the approximate subjective refraction in children after cycloplegia. METHODS 142 children aged 3 to 15 years were included. All children had their refractive status measured with the RMA-3000 autorefractometer (noncycloplegic autorefraction [AR]). Subsequently all children underwent cycloplegia and the refractive status was estimated again with the autorefractometer (cycloplegic autorefraction [ARC]) and traditional retinoscopy (RC) by examiners who were unaware of the results from the other techniques. RESULTS From 69 right eyes with negative sphere we observed that the sphere power was significantly higher (more than 0.5 diopters) in AR than in ARC (P = 0.0001) and RC (P = 0.0001). From the 73 normal and hyperopic right eyes we observed that the sphere power was significantly lower (more than 0.5 diopters) in AR than in ARC (P = 0.0001) and RC (P = 0.0001). CONCLUSIONS The use of the autorefractometer in children (in whom accommodation is more active than older patients) without cycloplegia may underestimate the actual hyperopia and overestimate the actual myopia. Manual retinoscopy is still the most accurate technique to estimate refractive status in children.
منابع مشابه
A comparison of cycloplegic and manifest refractions on the NR - 100 OF ( an objective Auto Refractometer )
The manifest (dry) and cycloplegic refractions of 50 eyes of 25 patients aged 8 to 28 years were studied on the Nikon Auto Refractometer NR-100OF (AR) and compared with the results of clinical refraction (CR) under homatropine and the final clinical acceptance on postmydriatic testing. Only patients in the younger age groups with low to moderate refractive errors were included in this study; hi...
متن کاملA comparison of cycloplegic and manifest refractions on the NR-1000F (an objective Auto Refractometer).
The manifest (dry) and cycloplegic refractions of 50 eyes of 25 patients aged 8 to 28 years were studied on the Nikon Auto Refractometer NR-1000F (AR) and compared with the results of clinical refraction (CR) under homatropine and the final clinical acceptance on postmydriatic testing. Only patients in the younger age groups with low to moderate refractive errors were included in this study; hi...
متن کاملThe influence of cycloplegia in objective refraction.
The purpose of this study was to compare refractions measured with an autorefractor and by retinoscopy with and without cycloplegia. The objective refractions were performed in 199 right eyes from 199 healthy young adults with a mean age of 21.6 +/- 2.66 years. The measurements were performed first without cycloplegia and repeated 30 min later with cycloplegia. Data were analysed using Fourier ...
متن کاملComparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children
PURPOSE To evaluate the performance of the hand-held and table-top autorefractokeratometer in measuring refractive errors by comparing them with cycloplegic retinoscopy. METHODS Included in the study were 112 eyes of 112 pediatric patients whose mean age was 6.78 ± 2.61 years (range, 2 to 12 years). The refractive errors of all the eyes were measured with and without cycloplegia using a hand ...
متن کاملCycloplegic refractions in healthy children aged 1 through 48 months.
OBJECTIVES To provide a description of refractive errors in healthy, term-born children, aged 1 through 48 months, and to test the hypotheses that spherical equivalent becomes significantly less hyperopic and less variable with increasing age. METHODS Following a prospective, cross-sectional design, cycloplegic retinoscopy was used to measure the refractive error in both eyes of 514 healthy, ...