Evaluate Shielding Design of the Brachytherapy Unit by Using Monte Carlo Simulation Code
نویسندگان
چکیده
Shielding design is necessary for brachytherapy treatment room in order to protect the general public and employees. The main objective of this study was to investigate whether the protective unit of our Brachytherapy Centre provided adequate protection to the health and safety assessment of radiobiological impact. In this study, we estimated the effect of radiobiological protection from a single Ir-192 brachytherapy source in Brachytherapy center by using MCNP5 Monte Carlo measurements. The room was based on the design specifications for the HDR 192 Ir treatment was modeled. The estimated dose rate range for HDR 192 Ir and public buildings is (0.45 0.64): (micro sivert) μSv/hour. Dose rates measured data for the current setup Brachytherapy HDR unit was approved and agreed quiet well with recommendation of International Atomic Energy Agency. The measured dose rate for public areas and controlled areas, compared with the reference value of 7.5 μSv/h and 0.5 μSv/h and concluded that we have enough shielding to the source but an over estimate with experimental measurements had been seen.
منابع مشابه
Calculating of Dose Distribution in Tongue Brachytherapy by Different Radioisotopes using Monte Carlo Simulation and Comparing by Experimental Data
Introduction: Among different kinds of oral cavity cancers, the frequency of tongue cancer occurrence is more significant. Brachytherapy is the most common method to cure tongue cancers. Long sources are used in different techniques of tongue brachytherapy. The objective of this study is to asses the dose distribution around long sources, comparing different radioisotopes as brachytherapy sourc...
متن کاملMonte Carlo Simulation for Treatment Planning Optimization of the COMS and USC Eye Plaques Using the MCNP4C Code
Introduction: Ophthalmic plaque radiotherapy using I-125 radioactive seeds in removable episcleral plaques is often used in management of ophthalmic tumors. Radioactive seeds are fixed in a gold bowl-shaped plaque and the plaque is sutured to the scleral surface corresponding to the base of the intraocular tumor. This treatment allows for a localized radiation dose delivery to the tumor with a ...
متن کاملA Monte Carlo Study on the Shielding Properties of a Novel Polyvinyl Alcohol (PVA)/WO3 Composite, Against Gamma Rays, Using the MCNPX Code
Background: In recent years, there has been an increased interest toward non-lead radiation shields consisting of small-sized filler particles doped into polymer matrices. In this paper, we study a new polyvinyl alcohol (PVA)/WO3 composite in the presence of high-energy gamma photons through simulation via the Monte Carlo N-Particle (MCNP) simulation code. Materia...
متن کاملCT-Based Brachytherapy Treatment Planning using Monte Carlo Simulation Aided by an Interface Software
Introduction: In brachytherapy, radioactive sources are placed close to the tumor, therefore, small changes in their positions can cause large changes in the dose distribution. This emphasizes the need for computerized treatment planning. The usual method for treatment planning of cervix brachytherapy uses conventional radiographs in the Manchester system. Nowadays, because of their advantages ...
متن کاملDetermination of Dosimetric characteristics of a New 192Ir-PDR Brachytherapy Source According to AAPM TG- 43 Protocol using Monte Carlo simulation technique
Introduction: 192Ir is one of the important sources frequently used in brachytherapy. Up to now, a lot of commercial models of this source have been made which Ir-192 has been recently added to them. The aim of the present study is to determine the dosimetric parameters of this new source model based on the recommendations of TG-43(U1) protocol using Monte Carlo simulation tech...
متن کامل