Late complications of transcatheter atrial septal defect closure requiring urgent surgery

نویسندگان

  • Pavle Kovacevic
  • Ilija Srdanovic
  • Vladimir Ivanovic
  • Jovan Rajic
  • Nemanja Petrovic
  • Lazar Velicki
چکیده

A 21-year-old female patient was transferred to our hospital after being diagnosed with massive pulmonary thromboembolism (PTE). Three years prior to admission, she underwent transcatheter closure of the secundum ASD with a 33 mm CardioSEAL-StarFLEX occluder (NMT Medical, Boston, MA, USA). A year ago, she was involved in a car accident and sustained significant blunt chest trauma. Transthoracic echocardiography confirmed the presence of thrombi in the right atrium and the pulmonary artery, with massive dilatation of the right ventricle and the pulmonary artery, along with severe pulmonary hypertension. Also, protrusion or dislodgement of the occluder was suspected. Her deteriorated clinical conditions warranted immediate surgery. The patient was put on a cardiopulmonary bypass (CPB) and the right atrium and the pulmonary artery were opened. Several thrombi were removed, the largest being 2 × 3 cm. The ASD occluder was identified with a thrombus attached to it and evident device-arm fracture (Figure 1). The occluder underwent almost complete healing with full endocardium covering except in the rim area. The device was removed and the ASD was repaired with a patch. Unfortunately, due to right heart failure, the patient could not be successfully weaned from the CPB, not even after an artificially created interatrial shunt, and she expired. Although one cannot say with absolute certainty that massive PTE developed because of device-related thrombosis, it seems intuitive that blood turbulence around the protruded umbrella and device-arm fracture could have acted as a nidus for repeated thrombus formation with subsequent embolization. The occluder malfunction (fracture) was most likely the result of sustained blunt chest trauma a year prior to admission. We hypothesize that the sudden increase in intrathoracic pressure during trauma as well as direct compression on the heart generated a point of high wall stress around the occluder’s septal insertion, which may have led to device fracture and dislodgment.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Urgent Surgical Intervention for Embolized Cardiac Occluder Devices: A Case Series

Introduction: In this study, we sought to illustrate our experience in urgent surgical management for embolized cardiac septal occlude devices resulting from trans-catheter closure of atrial septal defect and ventricular septal defect. Mathrials and Methods: We retrospectively reviewed four patients aged 2–10 years who underwent urgent surgery due to cardiac septal occluder embolization between...

متن کامل

Integrated Percutaneous Atrial Septal Defect Occlusion and Pulmonary Balloon Valvuloplasty

Introduction: Atrial Septal Defect (ASD) is one of the most common congenital heart diseases during childhood. Today, ASD closure is done by occlutech device via cardiac catheterization. ASD repair with transcatheter technique has shown high closure rate. However, coexistence of severe pulmonary stenosis and large defect in atrial septum is rare. When these two problems coexist, the result is i...

متن کامل

Perforation of the right atrium and the ascending aorta following percutaneous transcatheter atrial septal defect closure.

Transcatheter closure of secundum atrial septal defect has become an alternative to surgery. We present a patient with hemodynamic collapse secondary to cardiac perforation occurring 36 h after the placement of an Amplatzer Septal Occluder and discuss complications of this kind of device.

متن کامل

Congenital Heart Disease Transcatheter Closure of Atrial Septal Defects with the Amplatzer Septal Occluder Clinical Results

Objective: To assess the short-term outcome following transcatheter closure of secundum atrial septal defect (ASD) in our hospital. Patients and Methods: Between June 2000 and November 2001, 20 patients diagnosed as secundum atrial septal defects underwent transcatheter closure of their defects using the Amplatzer septal occluder (ASO) device. Indications for 18 patients were larger atrial sept...

متن کامل

Echocardiographic visualization of laceration of atrial septum during balloon sizing of atrial septal defect.

Transcatheter closure of atrial septal defect (ASD) has become an accepted alternative to surgery. A number of complications associated with ASD device closure have been recognized but most are rare or minor in severity. We report a rare complication of atrial septal laceration during transcatheter closure of secundum ASD. We discuss the diagnostic confusion, which resulted in the decision for ...

متن کامل

Emergent Surgical Intervention for Embolization of Atrial Septal Defect Closure Device

The percutaneous transcatheter closure of secundum atrial septal defect has recently become an increasingly widespread alternative to surgical closure in many centers. Although immediate, short, and intermediate term results of percutaneous transcatheter septal closure are promising, the procedure is not free from inherent complications that could be lethal. We report a case of device embolizat...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 13  شماره 

صفحات  -

تاریخ انتشار 2017