MRI-assisted PET motion correction for neurologic studies in an integrated MR-PET scanner.
نویسندگان
چکیده
UNLABELLED Head motion is difficult to avoid in long PET studies, degrading the image quality and offsetting the benefit of using a high-resolution scanner. As a potential solution in an integrated MR-PET scanner, the simultaneously acquired MRI data can be used for motion tracking. In this work, a novel algorithm for data processing and rigid-body motion correction (MC) for the MRI-compatible BrainPET prototype scanner is described, and proof-of-principle phantom and human studies are presented. METHODS To account for motion, the PET prompt and random coincidences and sensitivity data for postnormalization were processed in the line-of-response (LOR) space according to the MRI-derived motion estimates. The processing time on the standard BrainPET workstation is approximately 16 s for each motion estimate. After rebinning in the sinogram space, the motion corrected data were summed, and the PET volume was reconstructed using the attenuation and scatter sinograms in the reference position. The accuracy of the MC algorithm was first tested using a Hoffman phantom. Next, human volunteer studies were performed, and motion estimates were obtained using 2 high-temporal-resolution MRI-based motion-tracking techniques. RESULTS After accounting for the misalignment between the 2 scanners, perfectly coregistered MRI and PET volumes were reproducibly obtained. The MRI output gates inserted into the PET list-mode allow the temporal correlation of the 2 datasets within 0.2 ms. The Hoffman phantom volume reconstructed by processing the PET data in the LOR space was similar to the one obtained by processing the data using the standard methods and applying the MC in the image space, demonstrating the quantitative accuracy of the procedure. In human volunteer studies, motion estimates were obtained from echo planar imaging and cloverleaf navigator sequences every 3 s and 20 ms, respectively. Motion-deblurred PET images, with excellent delineation of specific brain structures, were obtained using these 2 MRI-based estimates. CONCLUSION An MRI-based MC algorithm was implemented for an integrated MR-PET scanner. High-temporal-resolution MRI-derived motion estimates (obtained while simultaneously acquiring anatomic or functional MRI data) can be used for PET MC. An MRI-based MC method has the potential to improve PET image quality, increasing its reliability, reproducibility, and quantitative accuracy, and to benefit many neurologic applications.
منابع مشابه
Improved PET Data Quantification in an Integrated Brain MR-PET Scanner
INTRODUCTION Simultaneous MR-PET data acquisition immediately brings to mind the possibility of improving the performance of one instrument by using the information obtained from the other modality. For example, a number of aspects have to be considered in PET for obtaining a correct quantitative measure of the activity concentration in a specific voxel. First, correcting for photon attenuation...
متن کاملTowards Lissajous navigator-based motion correction for MR-PET
Introduction With the combination of Magnetic Resonance Imaging (MRI) and Positron Emission Tomography (PET) into a single combined system, a novel imaging modality has become available. Exciting new study designs investigating different aspects of the exact same biochemical process are finally possible thanks to simultaneous MRI and PET acquisition. Due to long acquisition times of up to hours...
متن کاملFirst clinical experience with integrated whole-body PET/MR: comparison to PET/CT in patients with oncologic diagnoses.
UNLABELLED The recently introduced first integrated whole-body PET/MR scanner allows simultaneous acquisition of PET and MRI data in humans and, thus, may offer new opportunities, particularly regarding diagnostics in oncology. This scanner features major technologic differences from conventional PET/CT devices, including the replacement of photomultipliers with avalanche photodiodes and the ne...
متن کاملRespiratory motion correction of PET using MR-constrained PET-PET registration
BACKGROUND Respiratory motion in positron emission tomography (PET) is an unavoidable source of error in the measurement of tracer uptake, lesion position and lesion size. The introduction of PET-MR dual modality scanners opens a new avenue for addressing this issue. Motion models offer a way to estimate motion using a reduced number of parameters. This can be beneficial for estimating motion f...
متن کاملPET motion correction using MR-derived motion parameters
Medical Imaging Research Center, Department of Nuclear Medicine, KU Leuven, Belgium With the improving resolution of modern PET scanners, any slight motion during the scan can cause significant blurring and loss of resolution. MRI scanners have the capacity to perform quick successive scans and thus provide a means to track motion during a scan. Hence, with the advent of simultaneous PET-MR sca...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of nuclear medicine : official publication, Society of Nuclear Medicine
دوره 52 1 شماره
صفحات -
تاریخ انتشار 2011