Calibration of some Wright peak flow meters.

نویسندگان

  • J Fisher
  • A Shaw
چکیده

The performance of Wright peak flow meters is described using an absolute and precise calibration method. This showed errors as great as 25% with the peak flow meter and 40% with the mini peak flow meter. At large flow rates errors were less. The expected variation between individual peak flow meters was about 12%. The expected variation between individual mini peak flow meters was between 16% and 22%, depending on flow rate.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Portable peak flow meters: physical characteristics, influence of temperature, altitude, and humidity.

Little is known about the linearity of portable peak flow meters, or about physical gas factors affecting peak expiratory flow (PEF) readings. We therefore tested five portable peak flow meters of three types in an altitude chamber (sea level to 5,500 m) and in a climate chamber at sea level (7-37 degrees C) to determine the influence of the physical conditions of the gas on the reading of the ...

متن کامل

Re. Calibration of peak expiratory flow (PEF) measuring devices.

There are increasing numbers of devices coming onto the market for measuring PEP and of papers assessing their performance. In these papers a number of different ways of calibrating the devices have been used, as discussed in your correspondence columns [1]. What is the best way to calibrate a peak flow (PP) meter for clinical use? It seems not to be realized that compared with a large pneumota...

متن کامل

Bovine TB in badgers: a reappraisal of aetiology and pathogenesis.

I read with great interest the article of Miles et al. (1) about the performance of mini Wright peak flow meters after prolonged use. We performed a study (2) comparing the reliability of five new mini Wright peak flow meters before and after 10 000 compressed air impulses of about 4.50 1 min ’ flow, controlled by an electronically driven magnetic valve. There were no significant deviations of ...

متن کامل

Mini-Wright peak flow meters are reliable after 5 years use. Dutch CNSLD Study Group.

In studies with a long-term follow-up, peak expiratory flow (PEF) meters are often used to assess bronchial obstruction. The question arises whether data obtained with these frequently used meters are still reliable after several years of use, and whether the old meters should be renewed after a certain period. In the present study, we tested the reliability of PEF values measured with mini-Wri...

متن کامل

Eleven peak flow meters: a clinical evaluation.

Peak flow meters are essential tools in the management of asthma. Many types are on the market. A computer-driven piston pump is normally not available for evaluation of the various meters. Comparison with values from a pneumotachograph is an accepted way of testing peak flow meters. This study aims at comparing 11 peak flow meters, for accuracy and linearity. Seven adult peak flow meters were ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • British journal of anaesthesia

دوره 52 4  شماره 

صفحات  -

تاریخ انتشار 1980