Risk factors for failure of the 36 mm metal-on-metal Pinnacle total hip arthroplasty system

نویسندگان

  • G. S. Matharu
  • R. S. Nandra
  • F. Berryman
  • A. Judge
  • P. B. Pynsent
  • D. J. Dunlop
چکیده

AIMS To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure. PATIENTS AND METHODS We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression. RESULTS Failure occurred in 97 hips (17.0%). The ten-year cumulative failure rate was 27.1% (95% confidence interval (CI) 21.6 to 33.7). Primary implantation from 2006 onwards (hazard ratio (HR) 4.30; 95% CI 1.82 to 10.1; p = 0.001) and bilateral MoM hip arthroplasty (HR 1.59; 95% CI 1.03 to 2.46; p = 0.037) predicted failure. The effect of implantation year on failure varied over time. From four years onwards following surgery, hips implanted since 2006 had significantly higher failure rates (eight years 28.3%; 95% CI 23.1 to 34.5) compared with hips implanted before 2006 (eight years 6.3%; 95% CI 2.4 to 15.8) (HR 15.2; 95% CI 2.11 to 110.4; p = 0.007). CONCLUSION We observed that 36 mm MoM Pinnacle THAs have an unacceptably high ten-year failure rate, especially if implanted from 2006 onwards or in bilateral MoM hip patients. Our findings regarding implantation year and failure support recent concerns about the device manufacturing process. We recommend all patients undergoing implantation since 2006 and those with bilateral MoM hips undergo regular investigation, regardless of symptoms. Cite this article: Bone Joint J 2017;99-B:592-600.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Risk factors for failure of the 36 mm metal- on-metal Pinnacle total hip arthroplasty system A RETROSPECTIVE SINGLE-CENTRE COHORT STUDY

Patients and Methods We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multi...

متن کامل

Compression of the Femoral Vessels by a Pseudotumor after Metal-on-Metal Total Hip Arthroplasty

Here we present a case of pseudotumor following total hip arthroplasty (THA) that resulted in a circulatory disturbance caused by compression of the femoral vasculature. A 63-year-old man presented with pain, swelling, and redness of the left leg 5 years after primary metal-on-metal THA using the AML-Plus stem, Pinnacle® acetabular cup, and 36 mm diameter Ultamet™ metal head system (DePuy Ortho...

متن کامل

Fatal Cobalt Cardiomyopathy Following Revision Total Hip Arthroplasty – A Brief Report with Review of Literature

Ceramic bearing surfaces are being increasingly used in young patients undergoing total hip arthroplasty. However,failures have been reported including fractures even with the newer third generation ceramics. The recommendedtreatment for fracture of ceramic bearing surfaces is complete synovectomy and revision total hip arthroplasty.However, disappointing results have also bee...

متن کامل

Distributed analysis of hip implants using six national and regional registries: comparing metal-on-metal with metal-on-highly cross-linked polyethylene bearings in cementless total hip arthroplasty in young patients.

BACKGROUND The regulation of medical devices has attracted controversy recently because of problems related to metal-on-metal hip implants. There is growing evidence that metal-on-metal implants fail early and cause local and systemic complications. However, the failure associated with metal-on-metal head size is not consistently documented and needs to be communicated to patients and surgeons....

متن کامل

Retrospective cohort study of the performance of the Pinnacle metal on metal (MoM) total hip replacement: a single-centre investigation in combination with the findings of a national retrieval centre

OBJECTIVES To determine risk factors for revision in patients implanted with a commonly used metal on metal (MoM) hip replacement. DESIGN Retrospective cohort study in combination with a prospective national retrieval study (Northern Retrieval Registry (NRR)). SETTING Combined orthopaedic unit in combination with the NRR. PARTICIPANTS All patients implanted with a DePuy Pinnacle MoM hip p...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 99-B  شماره 

صفحات  -

تاریخ انتشار 2017