A Study to Compare the Radiation Absorbed Dose of the C-arm Fluoroscopic Modes
نویسندگان
چکیده
BACKGROUND Although many clinicians know about the reducing effects of the pulsed and low-dose modes for fluoroscopic radiation when performing interventional procedures, few studies have quantified the reduction of radiation-absorbed doses (RADs). The aim of this study is to compare how much the RADs from a fluoroscopy are reduced according to the C-arm fluoroscopic modes used. METHODS We measured the RADs in the C-arm fluoroscopic modes including 'conventional mode', 'pulsed mode', 'low-dose mode', and 'pulsed + low-dose mode'. Clinical imaging conditions were simulated using a lead apron instead of a patient. According to each mode, one experimenter radiographed the lead apron, which was on the table, consecutively 5 times on the AP views. We regarded this as one set and a total of 10 sets were done according to each mode. Cumulative exposure time, RADs, peak X-ray energy, and current, which were viewed on the monitor, were recorded. RESULTS Pulsed, low-dose, and pulsed + low-dose modes showed significantly decreased RADs by 32%, 57%, and 83% compared to the conventional mode. The mean cumulative exposure time was significantly lower in the pulsed and pulsed + low-dose modes than in the conventional mode. All modes had pretty much the same peak X-ray energy. The mean current was significantly lower in the low-dose and pulsed + low-dose modes than in the conventional mode. CONCLUSIONS The use of the pulsed and low-dose modes together significantly reduced the RADs compared to the conventional mode. Therefore, the proper use of the fluoroscopy and its C-arm modes will reduce the radiation exposure of patients and clinicians.
منابع مشابه
Evaluation of Radiation Exposure Pattern and Radiation Absorbed Dose Resulting from Occupational Exposure of Anesthesiologists to Ionizing Radiation
Introduction: Little information is available concerning the radiation exposure of anesthesiologists, and no such data have previously been collected in Iran. This prospective study was performed to determine the amount of radiation exposure of anesthesiologists for the purpose of assessing whether or not dangerous levels of radiation exposures were being reached, and to identify factors that c...
متن کاملAssessment of Patient Absorbed Radiation Dose during Hysterosalpingography: A Pilot Study in Southwest Nigeria
Background: Hysterosalpingography (HSG) is an indispensable tool for diagnosing infertility in females. The procedure exposes female reproductive organs to ionizing radiation as the genitals are irradiated during the process. Investigating patient absorbed dose during the procedures is essential for effective radiological protection of the patient. Objective: This study aims to investigat...
متن کاملInRaDoS: An internal radiation dosimetry computer program
Introduction: Internal radiation dosimetry is important from a radiation protection point of view and can help to optimize the radiation dose delivered to the workers, public, and patients. It has a rather simple protocol but needs a large amount of data. Therefore, it is difficult to do on a routine basis. The use of computer programs makes internal radiation dosimetry simpler...
متن کاملNanometric grafting of poly (N-isopropylacrylamide) onto polystyrene film by different doses of gamma radiation
Poly N-isopropylacrylamide was successfully grafted onto a polystyrene cell culture dish by gamma ray. In this study, the effect of a radiation dose (radiation absorbed dosages of 10, 20, 30, 40 KGy) under appropriate temperature and grafting conditions was investigated. The FTIR analysis showed the existence of the graft PNIPAAm on the substrate. The optimal value of the dose for grafting was ...
متن کاملBiological assessment and human absorbed dose estimation of [111In]In-DTPA-antiMUC1 as a radioimmunoconjugate for breast cancer imaging
Introduction: The aim of this study was to evaluate the human organ absorbed dose of [111In]In-DTPA-antiMUC1, as a newly developed radioimmunoconjugate. Methods: [111In]In-DTPA-antiMUC1 was prepared at optimized conditions while the radiochemical purity of the tracer was investigated using ITLC method. Biodistribution of the radiolab...
متن کامل