Reliability of MRSI brain temperature mapping at 1.5 and 3 T

نویسندگان

  • Michael J Thrippleton
  • Jehill Parikh
  • Bridget A Harris
  • Steven J Hammer
  • Scott I K Semple
  • Peter J D Andrews
  • Joanna M Wardlaw
  • Ian Marshall
چکیده

MRSI permits the non-invasive mapping of brain temperature in vivo, but information regarding its reliability is lacking. We obtained MRSI data from 31 healthy male volunteers [age range, 22-40 years; mean ± standard deviation (SD), 30.5 ± 5.0 years]. Eleven subjects (age range, 23-40 years; mean ± SD, 30.5 ± 5.2 years) were invited to receive four point-resolved spectroscopy MRSI scans on each of 3 days in both 1.5-T (TR/TE = 1000/144 ms) and 3-T (TR/TE = 1700/144 ms) clinical scanners; a further 20 subjects (age range, 22-40 years; mean ± SD, 30.5 ± 4.9 years) were scanned on a single occasion at 3 T. Data were fitted in the time domain to determine the water-N-acetylaspartate chemical shift difference, from which the temperature was estimated. Temperature data were analysed using a linear mixed effects model to determine variance components and systematic temperature changes during the scanning sessions. To characterise the effects of instrumental drift on apparent MRSI brain temperature, a temperature-controlled phantom was constructed and scanned on multiple occasions. Components of apparent in vivo temperature variability at 1.5 T/3 T caused by inter-subject (0.18/0.17 °C), inter-session (0.18/0.15 °C) and within-session (0.36/0.14 °C) effects, as well as voxel-to-voxel variation (0.59/0.54 °C), were determined. There was a brain cooling effect during in vivo MRSI of 0.10 °C [95% confidence interval (CI): -0.110, -0.094 °C; p < 0.001] and 0.051 °C (95% CI: -0.054, -0.048 °C; p < 0.001) per scan at 1.5 T and 3 T, respectively, whereas phantom measurements revealed minimal drift in apparent MRSI temperature relative to fibre-optic temperature measurements. The mean brain temperature at 3 T was weakly associated with aural (R = 0.55, p = 0.002) and oral (R = 0.62, p < 0.001) measurements of head temperature. In conclusion, the variability associated with MRSI brain temperature mapping was quantified. Repeatability was somewhat higher at 3 T than at 1.5 T, although subtle spatial and temporal variations in apparent temperature were demonstrated at both field strengths. Such data should assist in the efficient design of future clinical studies.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Measurement of human brain temperature changes during cooling

Introduction Therapeutic brain hypothermia is used with increasing frequency as a tool to mitigate neurologic injury [1]. It has been shown that moderate cooling of the brain after hypoxia-ischemia can reduce damage and improve functional outcome [2, 3]. Recently a new method for selective brain cooling was described and prototype device was tested on the pigs [4, 5]. Brain temperature decreasi...

متن کامل

Considerations in applying 3D PRESS H-1 brain MRSI with an eight-channel phased-array coil at 3 T.

The purpose of this study was to assess the benefits of a 3 T scanner and an eight-channel phased-array head coil for acquiring three-dimensional PRESS (Point REsolved Spectral Selection) proton (H-1) magnetic resonance spectroscopic imaging (MRSI) data from the brains of volunteers and patients with brain tumors relative to previous studies that used a 1.5 T scanner and a quadrature head coil....

متن کامل

DMSO as a potential contrast agent for brain tumours

MATERIALS AND METHODS: Nine C57BL/6 female mice were used in this study, 3 wild-type (WT) mice and six glioma-bearing mice induced as described in [4]. MRI studies were carried out at 7T in a horizontal magnet, anesthetizing the animals with isoflurane (1.5-2.0 %), monitoring their breathing pattern and controlling their rectal temperature (37oC). Initially, TMZ was diluted in DMSO and administ...

متن کامل

Spatial variability and reproducibility of GABA‐edited MEGA‐LASER 3D‐MRSI in the brain at 3 T

The reproducibility of gamma-aminobutyric acid (GABA) quantification results, obtained with MRSI, was determined on a 3 T MR scanner in healthy adults. In this study, a spiral-encoded, GABA-edited, MEGA-LASER MRSI sequence with real-time motion-scanner-instability corrections was applied for robust 3D mapping of neurotransmitters in the brain. In particular, the GABA+ (i.e. GABA plus macromolec...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 27  شماره 

صفحات  -

تاریخ انتشار 2014