A parametric study of cylindrical pedicle screw design implications on the pullout performance using an experimentally validated finite-element model.
نویسندگان
چکیده
The present study aims to the design of a finite-element model simulating accurately the pullout behaviour of cylindrical pedicle screws and predicting their pullout force. Three commercial pedicle screws, subjected to pure pullout from synthetic bone, were studied experimentally. The results were used for the design, calibration and validation of a finite-element model. Special attention was paid to the accurate simulation of the failure inside the host material under shear. For this purpose, a bilinear cohesive zone material model was adopted, controlling the mode-II debonding of neighbouring elements in the vicinity of the screw. Comparison between experimental and numerical results proved that the implementation of this approach can significantly enhance the accuracy of the numerical simulation of a screw's mechanical behaviour under pure pullout loads. The numerical model was used for the parametric study of various factors affecting the pullout performance of a cylindrical pedicle screw. It was concluded that the major parameter influencing the pullout force is the outer radius (increasing its value by 36% increases the pullout force by 34%). The influence of the purchase length of the screw is of similar quantitative nature. The respective dependence on the thread inclination, depth and pitch was significantly weaker.
منابع مشابه
Comparison and prediction of pullout strength of conical and cylindrical pedicle screws within synthetic bone
BACKGROUND This study was designed to derive the theoretical formulae to predict the pullout strength of pedicle screws with an inconstant outer and/or inner diameter distribution (conical screws). For the transpedicular fixation, one of the failure modes is the screw loosening from the vertebral bone. Hence, various kinds of pedicle screws have been evaluated to measure the pullout strength us...
متن کاملDependence of the pullout behaviour of pedicle screws on the screw-hosting material relative deformability
Indications exist in literature that a screw's pullout behaviour is among others influenced by the relative deformability of the screw and its hosting material. In addition it is known that the stress field developed in the vicinity of an orthopaedic implant significantly influences bone remodelling. In this context an experimentally validated finite element model of a screw and its hosting mat...
متن کاملShape Memory Alloy Expandable Pedicle Screw to Enhance Fixation in Osteoporotic Bone: Primary Design and Finite Element Simulation
The properties of shape memory alloys, specifically the equiatomic intermetallic NiTi, are unique and significant in that they offer simple and effective solutions for some of the biomechanical issues encountered in orthopedics. Pedicle screws, used as an anchoring point for the implantation of spinal instrumentations in the spinal fracture and deformity treatments, entail the major drawback of...
متن کاملMultiobjective Optimization Design of Spinal Pedicle Screws Using Neural Networks and Genetic Algorithm: Mathematical Models and Mechanical Validation
Short-segment instrumentation for spine fractures is threatened by relatively high failure rates. Failure of the spinal pedicle screws including breakage and loosening may jeopardize the fixation integrity and lead to treatment failure. Two important design objectives, bending strength and pullout strength, may conflict with each other and warrant a multiobjective optimization study. In the pre...
متن کاملEffect of bone material properties on effective region in screw-bone model: an experimental and finite element study
BACKGROUND There have been numerous studies conducted to investigate the pullout force of pedicle screws in bone with different material properties. However, fewer studies have investigated the region of effect (RoE), stress distribution and contour pattern of the cancellous bone surrounding the pedicle screw. METHODS Screw pullout experiments were performed from two different foams and the c...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Medical engineering & physics
دوره 32 2 شماره