Increased total knee arthroplasty ultra-high molecular weight polyethylene wear using a clinically relevant hyaluronic acid simulator lubricant.

نویسندگان

  • J DesJardins
  • A Aurora
  • S L Tanner
  • T B Pace
  • K B Acampora
  • M Laberge
چکیده

In this study, osteoarthritic and periprosthetic synovial fluid samples were rheologically and biochemically compared to develop a hyaluronic acid (HA) supplemented bovine serum (BS) lubricant that mimicked the properties of human joint synovial fluid. The effect of this BS + HA lubricant (50 per cent bovine calf serum + 1.5 g/l HA) on the wear rate of ultra-high molecular weight polyethylene (UHMWPE) during a total knee replacement wear test was then investigated. In conjunction with biochemical similarities, the rheological analysis showed that the BS + HA lubricant viscosity was not statistically different to aspirated total knee arthroplasty (TKA) revision joint fluid viscosity over a range of physiologic shear rates. Gravimetric results at 5 million wear testing cycles showed that the BS + HA lubricant produced an average of 6.88 times more UHMWPE wear than 50 per cent bovine serum lubricant alone. The BS + HA lubricated CoCr femoral component surfaces revealed pitting and surface roughening that was not observed using standard bovine serum only lubricants, but that was similar to the metallic surface corrosion observed on in vivo CoCr femoral component retrievals. These findings support the hypothesis that the addition of HA to simulator lubricant is capable of producing CoCr femoral component surface damage similar to that observed in vivo.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of lubricant composition on the fatigue properties of ultra-high molecular weight polyethylene for total knee replacement.

Ultrahigh molecular weight polyethylene (UHMWPE) fatigue is a critical factor affecting the longevity of total knee replacement (TKR) bearings. With the increased need for laboratory studies to mimic near in vivo conditions for accurate characterization of material performance, the present study investigated the role of hyaluronic acid (HA) in testing lubricant on the crack growth response of U...

متن کامل

Albumin Protein Cleavage Affects the Wear and Friction of Ultra-High Molecular Weight Polyethylene

It is well established that the total protein concentration and albumin-to-globulin ratio influence the wear of ultra-high molecular polyethylene (UHMWPE, "polyethylene") in joint prostheses. A factor on wear not yet studied, but of possible clinical relevance, is protein cleavage. Such cleavage is expected in the presence of an inflammatory response and as a result of wear processes at the art...

متن کامل

Reduction of Wear in Fixed Bearing Total Knee Replacement Using Crosslinked Uhmwpe

Introduction Reduction of ultra high molecular weight polyethylene (UHMWPE) surface wear in total knee replacement (TKR) bearings may delay the onset of osteolysis and subsequent loosening of components. Crosslinking has been shown to increase the wear resistance of UHMWPE bearings in multidirectional total hip replacement applications when articulating against smooth counterfaces [1]. The aim ...

متن کامل

Generation of a large volume of clinically relevant nanometre-sized ultra-high-molecular-weight polyethylene wear particles for cell culture studies

It has recently been shown that the wear of ultra-high-molecular-weight polyethylene in hip and knee prostheses leads to the generation of nanometre-sized particles, in addition to micron-sized particles. The biological activity of nanometre-sized ultra-high-molecular-weight polyethylene wear particles has not, however, previously been studied due to difficulties in generating sufficient volume...

متن کامل

Knee Simulator Wear of Cross-linked Uhmwpe with Normal and Mal-rotation Kinematics

Introduction: Wear of ultra-high molecular weight polyethylene (UHMWPE) is a major factor that affects longevity of the total joint replacement. In total hips (THR), cross-linking of polyethylene has been shown to be highly effective in reducing wear both clinically and experimentally [1-3]. Also, in TKR, Schmidig et al. [4] showed a 90% reduction of wear rate in 10 Mrad compared to 3 Mrad tibi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine

دوره 220 5  شماره 

صفحات  -

تاریخ انتشار 2006