Evaluation of 3-D Printed Immobilisation Shells for Head and Neck IMRT
نویسندگان
چکیده
This paper presents the preclinical evaluation of a novel immobilization system for patients undergoing external beam radiation treatment of head and neck tumors. An immobilization mask is manufactured directly from a 3-D model, built using the CT data routinely acquired for treatment planning so there is no need to take plaster of Paris moulds. Research suggests that many patients find the mould room visit distressing and so rapid prototyping could potentially improve the overall patient experience. Evaluation of a computer model of the immobilization system using an anthropomorphic phantom shows that >99% of vertices are within a tolerance of ±0.2 mm. Hausdorff distance was used to analyze CT slices obtained by rescanning the phantom with a printed mask in position. These results show that for >80% of the slices the median “worse-case” tolerance is approximately 4 mm. These measurements suggest that printed masks can achieve similar levels of immobilization to those of systems currently in clinical use.
منابع مشابه
The effect of high energy photon beam in dose- volume parameters and integral dose of normal structures in Head and neck IMRT
Introduction: The aim of this study is to investigate the effect of partially used high photon energies on dosimetric parameters and integral doses of nasopharynx IMRT plans. Materials and Methods: Two 7 coplanar IMRT plan were created for eleven patients (6 MV only IMRT plan in both phase1 and phase2 and 6 MV phase1-mixed energies (6 MV and 15 MV) phase2...
متن کاملDose Distribution Analysis of Rapid Arc and Intensity Modulated Radiotherapy Plan in Head and Neck Cancer
Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nom...
متن کاملSlant board immobilisation of head-and-neck radiotherapy patients who cannot tolerate a flat position
Purpose: Patients treated with intensity-modulated radiation therapy (IMRT) for head-and-neck cancer are often positioned supine on a carbon fibre board to which a thermoplastic mask is attached to immobilise the head and shoulders. For patients unable to tolerate a supine position, we developed a tilting board that accommodates a full-scale head-and-shoulder mask. Materials and methods: Phanto...
متن کاملComparison of set up accuracy among three common immobilisation systems for intensity modulated radiotherapy of nasopharyngeal carcinoma patients
INTRODUCTION In intensity modulated radiotherapy (IMRT) of nasopharyngeal carcinoma (NPC) patients, an effective immobilisation system is important to minimise set up deviation. This study evaluated the effectiveness of three immobilisation systems by assessing their set up deviations. METHODS Patients were randomly assigned to one of the three immobilisation systems: (1) supine on head rest ...
متن کاملSetup errors in patients with head-neck cancer (HNC), treated using the Intensity Modulated Radiation Therapy (IMRT) technique: how it influences the customised immobilisation systems, patient’s pain and anxiety
BACKGROUND In patients with head-neck cancer treated with IMRT, immobility of the upper part of the body during radiation is maintained by means of customised immobilisation devices. The main purpose of this study was to determine how the procedures for preparation of customised immobilisation systems and the patients characteristics influence the extent of setup errors. METHODS A longitudina...
متن کامل