The Vestibular Implant: Hearing Preservation during Intralabyrinthine Electrode Insertion—A Case Report
نویسندگان
چکیده
OBJECTIVE The vestibular implant seems feasible as a clinically useful device in the near future. However, hearing preservation during intralabyrinthine implantation remains a challenge. It should be preserved to be able to treat patients with bilateral vestibulopathy and (partially) intact hearing. This case study investigated the feasibility of hearing preservation during the acute phase after electrode insertion in the semicircular canals. METHODS A 40-year-old woman with normal hearing underwent a translabyrinthine approach for a vestibular schwannoma Koos Grade IV. Hearing was monitored using auditory brainstem response audiometry (ABR). ABR signals were recorded synchronously to video recordings of the surgery. Following the principles of soft surgery, a conventional dummy electrode was inserted in the lateral semicircular canal for several minutes and subsequently removed. The same procedure was then applied for the posterior canal. Finally, the labyrinthectomy was completed, and the schwannoma was removed. RESULTS Surgery was performed without complications. No leakage of endolymph and no significant reduction of ABR response were observed during insertion and after removal of the electrodes from the semicircular canals, indicting no damage to the peripheral auditory function. The ABR response significantly changed when the semicircular canals were completely opened during the labyrinthectomy. This was indicated by a change in the morphology and latency of peak V of the ABR signal. CONCLUSION Electrode insertion in the semicircular canals is possible without acutely damaging the peripheral auditory function measured with ABR, as shown in this proof-of-principle clinical investigation.
منابع مشابه
Achievement of hearing preservation in the presence of an electrode covering the residual hearing region
CONCLUSIONS With full insertion with a long electrode, hearing preservation can be achieved even in the presence of a long electrode covering the residual hearing region. OBJECTIVES Advances in developing new atraumatic concepts of electrode design as well as surgical technique have enabled hearing preservation after cochlear implantation surgery, and EAS (electric acoustic stimulation) accom...
متن کاملIntralabyrinthine Penetrating Ventilation Tube with Preservation of Hearing: An Unusual Clinical Situation
Introduction Traumatic perilymphatic fistula is not a rare event with regards to sport activities or traffic accident. However, iatrogenic damage to the inner ear can occur following the common use of grommets and ventilation tube insertion. Objectives To report an unusual case of insertion of aeration tube into the vestibule trough the stapes footplate. Resumed Report A 62-year-old woman exper...
متن کاملInvestigating the Effect of Cochlear Size in Insertion of Electrode Depth in Patients with Cochlear Implantation evaluated by CT-Scan
Background Cochlear implant surgery is an invasive procedure for patients with bilateral Sensorineural hearing loss and may cause some risks such as cochlear damage. We aimed to evaluate the effectiveness of cochlear measures obtained by computed tomography (CT) scan in predicting depth of cochlear implant insertion. Materials and Methods This study was conducted in a retrospective and cross-se...
متن کاملAn overview of cochlear implant electrode array designs
Cochlear implant electrode arrays are designed with specific characteristics that allow for the preservation of intra-cochlear structures during the insertion process, as well as during explantation. Straight lateral wall (LW) electrode arrays and pre-curved modiolar hugging (MH) electrode arrays are the two types that are commercially available. Although there is a third type of electrode arra...
متن کاملRound window electrocochleography before and after cochlear implant electrode insertion.
OBJECTIVE/HYPOTHESIS Previous reports have documented the feasibility of utilizing electrocochleographic (ECoG) responses to acoustic signals to assess trauma caused during cochlear implantation. The hypothesis is that intraoperative round window ECoG before and after electrode insertion will help predict postoperative hearing preservation outcomes in cochlear implant recipients. STUDY DESIGN...
متن کامل