Technical Note : Characterization of a static thermal - gradient CCN counter

نویسندگان

  • G. P. Frank
  • U. Dusek
  • M. O. Andreae
چکیده

The static (parallel-plate thermal-gradient) diffusion chamber (SDC) was one of the first instruments designed to measure cloud condensation nuclei (CCN) concentrations as a function of supersaturation. It has probably also been the most widely used type of CCN counter. This paper describes the detailed experimental characterization of a SDC CCN counter, including calibration with respect to su-persaturation and particle number concentration. In addition, we investigated the proposed effect of lowered supersatura-tion because of water vapor depletion with increasing particle concentration. The results obtained give a better understanding why and in which way it is necessary to calibrate the SDC CCN counter. The calibration method is described in detail and can, in parts, be used for calibrations also for other types of CCN counters. We conclude the following: 1) it is important to experimentally calibrate SDC CCN counters with respect to su-persaturation, and not only base the supersaturation on the theoretical description of the instrument; 2) the number concentration calibration needs to be performed as a function of supersaturation, also for SDC CCN counter using the photographic technique; and 3) we observed no evidence that water vapor depletion lowered the supersaturation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Characterization of a static thermal-gradient CCN counter

The static (parallel-plate thermal-gradient) diffusion chamber (SDC) was one of the first instruments designed to measure cloud condensation nuclei (CCN) concentrations as a function of supersaturation. It has probably also been the most widely used type of CCN counter. This paper describes the detailed experimental characterization of 5 a SDC CCN counter, including calibration with respect to ...

متن کامل

Evaluating the Effects of Ceramic Layer and Thermal Dam on Optimizing the Temperature Gradient of a Gasoline Engine Piston (TECHNICAL NOTE)

The purpose of this paper is to evaluate the effect of different methods for improving the temperature gradient of a specified gasoline engine piston. With a robust FE based software, 3D thermal analyses have been carried out for the piston model. Unlike previous studies, the effects of both fully and locally ceramic layers on the crown top surface were considered. It was found that a fully cer...

متن کامل

Mapping the Operation of the DMT Continuous Flow CCN Counter

This work thoroughly analyzes a new commercial instrument for measuring Cloud Condensation Nuclei (CCN), the Droplet Measurement Technologies Cylindrical Continuous-Flow Streamwise Thermal Gradient CCN Chamber (CFSTGC). This instrument can measure CCN concentrations at supersaturations from 0.06% to 3% (potentially up to 6%), at a 1 Hz sampling rate that is sufficient for airborne operation. Ou...

متن کامل

Improvement of the Solvent Extraction of Rhenium from Molybdenite Roasting Dust Leaching Solution using Counter-current Extraction by a Mixer-settler Extractor (TECHNICAL NOTE)

Continuous counter-current extraction of rhenium from roasting dust leach liquor was carried out using a mixer-settler extractor. Tributylphosphate was used as the extractant diluted in kerosene. The effects of the flow rates and extraction stages number were investigated. The extraction efficiency was affected by the flow rates of the aqueous and organic phases, and its mechanism was qualitati...

متن کامل

Scanning Flow CCN Analysis - A Method for Fast Measurements of CCN Spectra

The Continuous-Flow Streamwise Thermal-Gradient Cloud Condensation Nuclei Counter is a commercially-available instrument that is widely used for laboratory and field measurements of cloud condensation nuclei (CCN). Typically, the instrument is operated at a constant flow rate, and supersaturation is adjusted by changing the column streamwise temperature difference, which is slow and may introdu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2007