Biomechanical testing of spinal segment fixed by thoracolumbar spine locking plate on the swine lumbar spine.
نویسندگان
چکیده
BACKGROUND The aim of the experiment was to compare the mechanical properties of intact spinal segment with impaired intervertebral disc and impaired intervertebral disc fixed by TSLP (Thoracolumbar Spine Locking Plate). METHODS AND RESULTS Spinal specimens were taken from domestic swine. A total of 8 test mechanical states (intact, impaired and fixed) were modeled and the mechanical properties, expressed by the value of moment of couple necessary to twist the specimen at tensile force F = 200 N and the value of moments necessary for extension straining, were determined. The study was based on in vitro biomechanical testing of the TSLP plate used to stabilize the front thoracolumbar column of spinal segments taken from a pig. The plate was used for monosegmental fixation. The disc was cut by scalpel to simulate the Type A injury to front spinal column. In each state (intact, impaired or fixed), specimens were subjected to a tension load of prescribed force and, then, twisted by a given angle. Subsequently, extension load of intact, impaired and impaired & fixed segment was measured. Statistical evaluation verified the hypothesis of the different behavior of intact, impaired and fixed specimens - both for tension & torsion load and extension load. The analyses did not indicate different mechanical behavior of intact and fixed specimens. In other words, monosegmental fixation of both impaired and intact specimens by TSLP Synthes implant will lead to similar mechanical behavior of these specimens. Further, we found that intact and fixed specimens show non-symmetric behavior at positive and negative twisting angles. This was not observed for impaired specimens. CONCLUSION Several stabilization systems were developed to stabilize the front thoracolumbar spinal column. Surgery of the anterior column of injured spine should restore the correct position of the spine, ensure decompression of vertebral canal when neural structures are compressed, and stabilize the spine to allow immediate loading and mobilization of the patient. The aim of this study was to compare mechanical properties of intact spinal segment, impaired spinal segment and impaired spinal segment stabilized by TSLP Synthes implant. The problems were solved by experimental modeling using a testing machine that simulated loads for several mechanical states of the spinal segment. Favorable mechanical properties of TSLP Synthes fixator were demonstrated. The experimental results will be used for subsequent computational modeling of the spinal segment in all experimentally solved states.
منابع مشابه
Two in vivo surgical approaches for lumbar corpectomy using allograft and a metallic implant: a controlled clinical and biomechanical study.
BACKGROUND CONTEXT Both bone graft and metallic implants have been used in combination with the necessary anterior rod or plate instrumentation to fill the voids left by vertebral body removal, with the ultimate goal of restoring stability. One type of device that has recently been introduced is an expandable titanium telescoping cage that is designed to be used as a strut implant to fill corpe...
متن کاملBiomechanical Testing of Spinal Segment Fixed by Arcofix System on the Swine Spine
STUDY DESIGN An in vitro biomechanical study. PURPOSE To evaluate the mechanical properties of the spinal segment in the intact, injured, and stabilized state after fixation by an Arcofix implant. OVERVIEW OF LITERATURE Several types of thoracolumbar spine injury necessitates anterior instrumentation. The Arcofix plate represents the latest generation of angular stablity systems. The biomec...
متن کاملEstimation of spinal loads using a detailed finite element model of the L4-L5 lumbar segment derived by medical imaging kinematics; a feasibility study
Introduction: Low back pain is the most prevalent orthopedic disorder and the first main cause of poor working functionality in developed as wells as many developing countries. In Absence of noninvasive in vivo measurement approaches, biomechanical models are used to estimate mechanical loads on human joints during physical activities. To estimate joint loads via musculoskelet...
متن کاملThoracolumbar and Lumbar Burst Fractures
Approximately 79,000 spinal fractures occur annually in the United States. Roughly three out of every four fractures involve either the thoracic or the lumbar spine1,2. The most common site of injury in the thoracic and lumbar spine is the thoracolumbar junction, the mechanical transition zone between the relatively rigid thoracic spine and the more flexible lumbar region3-5. The thoracolumbar ...
متن کاملProduction of Reprocucible Specific Spinal Fractures in Calf Thoracolumbar Spine Specimens
Introduction: In vitro biomechanical testing of the spine is a well established method to analyze, compare and evaluate different strategies in treatment of spinal fractures [1-3]. Fracture morphology has an important influence on the clinical decision making process of its treatment. By investigating spinal fracture care in vitro, the different types of fractures leading to distinct treatment ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Biomedical papers of the Medical Faculty of the University Palacky, Olomouc, Czechoslovakia
دوره 154 4 شماره
صفحات -
تاریخ انتشار 2010