Rapid and simultaneous determination of 25-OH-vitamin D2 and D3 in human serum by LC/MS/MS: Validation and uncertainty assessment
نویسندگان
چکیده
25-OH Vitamin D3 and 25-OH Vitamin D2 were determined by rapid, simple and validated LC/MS/MS method without using SPE and isotopic standard. The relative standard deviations (RSD) were found to be 3.1 and 2.8 % for 25-OH Vitamin D3 and 25-OH Vitamin D2, respectively. The correlation coefficient was grater than 0.99 for both analytes in the calibration range. The recoveries at 100 μg L concentration averaged to 92 % for 25-OH vitamin D3 and 94 % for 25-OH-vitamin D2. The uncertainty of the measurements for 25-OH vitamin D3 and for 25-OH-vitamin D2 was 7.77 % and 7.24 %, respectively.
منابع مشابه
A fast and simple method for simultaneous measurements of 25(OH)D, 24,25(OH)2D and the Vitamin D Metabolite Ratio (VMR) in serum samples by LC-MS/MS.
BACKGROUND Rapid, easy and reliable measurement of the major vitamin D metabolites is required in order to fulfill the needs of a clinical routine laboratory. To overcome these challenges, we have developed and validated a LC-MS/MS method for the quantification of 25-hydroxyvitamin D2 and D3, epi-25-hydroxyvitamin D3 and 24,25-dihydroxyvitamin D3. METHODS Sample preparation was based on preci...
متن کاملPerformance evaluation of two immunoassays for 25-hydroxyvitamin D
Although immunoassays in measuring 25-hydroxyvitamin D [25(OH)D] have been improved recently, relatively large differences are still seen between results of 25(OH)D measured by immunoassays and by liquid chromatography-tandem mass spectrometry (LC-MS/MS). In the present studies, we compared two immunoassays with LC-MS/MS for measuring 25(OH)D concentrations. Concentrations of 25-hydroxyvitamin ...
متن کاملCandidate reference measurement procedures for serum 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 by using isotope-dilution liquid chromatography-tandem mass spectrometry.
BACKGROUND 25-hydroxyvitamin D [25(OH)D] assays are characterized by poor between-assay comparability. This result emphasizes the need for reference measurement procedures (RMPs) to establish calibration traceability and assist in method validation. We aimed at developing candidate RMPs on the basis of isotope- dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) for separate q...
متن کاملRapid Analysis of Vitamin D in Serum Using Triple Quadrupole LC/MS
A rapid method for the quantification of 25-OH D2 and 25-OH D3 in serum was developed using the Agilent 1200 Series Rapid Resolution LC (RRLC) system coupled with an Agilent 6430 Triple Quadrupole LC/MS with multimode ionization (MMI). Solid phase extraction minimizes matrix interferences and ion suppression, while enabling automation. Excellent linearity of quantification (R2 > 0.99) was obtai...
متن کاملMethod evaluation study of a new generation of vitamin D assays
INTRODUCTION Recently several diagnostic manufacturers have launched new 25-hydroxy-vitamin D (25[OH]D) assays, which are aligned to the National Institute of Standards and Technology (NIST) Standard Reference Materials (SRM) (NIST, Gaithersburg, Maryland). The aim of this study was to compare the performance of one liquid chromatography-tandem mass spectrometry (LC-MS/MS) method, one enzyme li...