Evolving use of embolisation coils for occlusion of the arterial duct.

نویسندگان

  • E Rosenthal
  • S A Qureshi
  • J Reidy
  • E J Baker
  • M Tynan
چکیده

OBJECTIVE To assess the outcome of arterial duct occlusion with coils chosen according to the duct morphology. DESIGN Retrospective study. SETTING Paediatric cardiology centre. PATIENTS Coil occlusion was attempted in 57 patients aged 0.5 to 15 (median 3.7) years and weighing 5-59 (median 14) kg between January 1991 and December 1995. A residual leak was present in 8 patients after umbrella closure and in 4 patients after duct ligation. METHODS Coils of 4 different types were implanted through 4 or 5 F femoral artery catheters. Platinum or Interlocking Detachable 0.018 inch coils were deployed completely inside tubular ducts. Gianturco or PDA controlled release 0.038 inch coils were implanted to straddle short, post ligation and post umbrella ducts. RESULTS Coil implantation was successful in 54/57 patients. At 1 year the cumulative occlusion rate was 53/57 ducts (93%) on an intention to treat analysis. A single coil was implanted in 37 (69%), 2 coils in 10 (19%), 3 coils in 3 (5%) and 4 coils in 4 (7%) of the 54 successful procedures. Duct occlusion was documented at the end of the procedure in 31%, by the following day in 83%, by 6 weeks in 87%, by 6 months in 96%, and by 1 year in 98%. Coil embolisation occurred in 6/58 procedures (10%), with a 50% rate in the first year of implantation (1/2 patients) falling to 7% in the last year (3/42 patients). All embolised coils were easily retrieved. CONCLUSIONS Occlusion of small to moderate size arterial ducts, including residual post umbrella or post ligation ducts, was readily accomplished by coils selected according to the duct anatomy. This has both cost and practical benefits.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Spring coil retraction in coil occlusion of persistent ductus arteriosus.

AIMS To present the short and intermediate term results of coil occlusion of persistent ductus arteriosus and the results of radiographic measurements of spring coils implanted to treat patent ducts. PATIENTS 22 children underwent coil occlusion. Their ages ranged from 2 years 9 months to 12 years 10 months (mean (SD) age, 6.5 (3.6) years). The duct diameter ranged from 1.0 to 3.5 mm at the n...

متن کامل

Early implantation of multiple spring coils for severe haemolysis after incomplete transcatheter occlusion of persistent arterial duct.

An 18 month old girl with an angiographically measured ductus of 4.5 mm underwent transcatheter occlusion of the persistent arterial duct with a 17 mm Rashkind umbrella and an occluding spring coil. Severe intravascular haemolysis developed 20 hours later. Significant residual ductal leakage was noted and the residual duct measured 6 mm. Previous underestimation might have been related to ducta...

متن کامل

Lung perfusion studies after detachable coil occlusion of persistent arterial duct.

OBJECTIVE To evaluate relative lung perfusion following complete occlusion of persistent arterial duct with detachable Cook coils. METHODS Ductal occlusion using detachable coils was performed in 35 patients (median age 3.9 years, range 0.5 to 16; 32 native ducts, three patients with previous devices). If the duct could be crossed with a 0.035 inch guidewire and a 4 F catheter after coil impl...

متن کامل

Umbrella occlusion of persistent arterial duct in children under two years.

OBJECTIVE To assess the use of trans-catheter occlusion of a persistent arterial duct in symptomatic children < 2 years of age. DESIGN Descriptive study of selected, non-randomised infants with persistent arterial duct who underwent attempted umbrella occlusion. SETTING Tertiary referral centre. PATIENTS Between June 1990 and April 1993, 29 young children with a symptomatic persistent art...

متن کامل

Haemolysis following implantation of duct occlusion coils.

OBJECTIVES To describe the incidence and management of haemolysis after transcatheter coil occlusion of the arterial duct. DESIGN Prospective clinical and echocardiographic follow up of patients who have undergone implantation of the Cook detachable duct occlusion coil. SETTING Tertiary paediatric cardiac centre. PATIENTS Five cases of haemolysis (two girls aged 6 and 11 months; three boy...