Pacemaker sensitivity to 50 Hz noise voltages.

نویسندگان

  • Werner Irnich
  • Mary Kay Steen-Mueller
چکیده

AIMS European standards specify that pacemakers (PM) should be resistant to electromagnetic interference (EMI) up to an upper borderline voltage as a function of frequency. Electromagnetic interference fields should either remain below this upper borderline voltage level or be identified and isolated from the general population. Physicians caring for PM patients need to be aware of potential problems relating to EMI. For example, sensitivity should be programmed to avoid sensing EMI below the recommended borderline voltage level. The susceptibility of a pacemaker (PM) to 50 Hz noise is an important parameter of EMI and depends on the programmed sensitivity [sensitivity setting (SS)]. We studied SS and 50 Hz noise thresholds in a large population and determined the borderline SS, defined as the SS below which 50 Hz noise was sensed. Our results should be taken into consideration in programming the SS to protect patients from the adverse effects of EMI. METHODS AND RESULTS Measurements were performed on 189 PMs explanted after death. All PMs studied were implanted in 1998 or later. Sensitivity setting and sensing configuration (unipolar or bipolar) were left as programmed during lifetime. The ventricular SS and 50 Hz noise thresholds were measured according to the European pacemaker standard. The signal-to-noise ratios (S/N) were derived from the heart test and noise test signal thresholds. The S/N for pulsed 50 Hz noise of five manufacturers tested ranged from 0.435 to 0.59. The S/N for 50 Hz continuous noise for four manufactures other than Medtronic was higher, ranging from 0.458 to 0.623. No PM showed a ratio of 1 or better. Medtronic PMs reacted differently to 50 Hz continuous noise than the other brands. In 24 Medtronic PMs, the continuous noise threshold was evaluated with two heart test signal amplitudes: either 10 mV or threshold level. In tests at threshold amplitudes, voltages between 0.1 and 0.85 mV elicited interference proving that Medtronic PM reacted extremely sensitively to noise. At 10 mV heart test amplitude, the noise threshold was inversely proportional to the SS, i.e. higher SS resulted in lower noise thresholds. Noise immunity increases with increasing heart test signal amplitude. CONCLUSION All tested PMs reacted to pulsed 50 Hz waves as if they were heart signals and were inhibited. Continuous noise above noise threshold evoked asynchronous pacing at noise rate. All PMs had an S/N ratio <1, indicating that the heart signals were amplified less than noise. The European Standard requires that unipolarly sensing PMs tolerate noise up to 2 mV. However, an SS of 2 mV does not guarantee a noise tolerance of 2 mV. In order to fulfil this requirement, SS in the majority of PMs must be programmed >2 mV. In Medtronic PMs, the continuous noise threshold is paradoxical as it is higher with decreasing SS. As a good compromise in Medtronic PMs, SS should be ∼3 mV to guarantee sufficient protection from pulsed and continuous noise, assuming ventricular heart signals of 10 mV or more.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Computation of Pacemakers Immunity to 50 Hz Electric Field: Induced Voltages 10 Times Greater in Unipolar Than in Bipolar Detection Mode

Thisstudy aims to compute 50 Hz electric field interferences on pacemakers for diverse lead configurations and implantation positions. Induced phenomena in a surface-based virtual human model (standing male grounded with arms closed, 2 mm resolution) are computed for vertical exposure using CST EM® 3D software, with and without an implanted pacemaker. Induced interference voltages occurring on ...

متن کامل

Improvements in Electrocardiography Smoothening and Amplification

The electrocardiogram (ECG or EKG) is a graphic recording of the time-variant voltages produced by the myocardium during the cardiac cycle. The P, QRS, and T waves reflect the rhythmic electrical depolarization and re-polarization of the myocardium associated with the contractions of the atria and ventricles. The electrocardiogram is generally used clinically in diagnosing diseases of the heart...

متن کامل

Active medical implants and occupational safety--measurement and numerical calculation of interference voltage.

Low frequency electric and magnetic fields may interfere with implanted cardiac pacemakers causing a life-threatening malfunction of the device. In order to assess the safety of workers in the vicinity of industrial electrical devices the interference voltage at the input port of a pacemaker is an important measure. In order to investigate the coupling of fields emanating from electrical device...

متن کامل

Voltage-gated sodium channels improve contrast sensitivity of a retinal ganglion cell.

Voltage-gated channels in a retinal ganglion cell are necessary for spike generation. However, they also add noise to the graded potential and spike train of the ganglion cell, which may degrade its contrast sensitivity, and they may also amplify the graded potential signal. We studied the effect of blocking Na+ channels in a ganglion cell on its signal and noise amplitudes and its contrast sen...

متن کامل

A Phase Noise Reduction Technique in LC Cross-coupled Oscillators with Adjusting Transistors Operating Regions

In this paper, an intuitive analysis of a phase noise reduction technique is done, and then a modified structure is proposed to achieve higher phase noise reduction than the original one. This method reduces the impact of noise sources on the phase noise by decreasing closed-loop gain in zero-crossings points and moving this high closed-loop gain to the non-zero-crossings points. This reduction...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology

دوره 13 9  شماره 

صفحات  -

تاریخ انتشار 2011