The ‘Sphere’: A Dedicated Bifurcation Aneurysm Flow-Diverter Device
نویسندگان
چکیده
We present flow-based results from the early stage design cycle, based on computational modeling, of a prototype flow-diverter device, known as the 'Sphere', intended to treat bifurcation aneurysms of the cerebral vasculature. The device is available in a range of diameters and geometries and is constructed from a single loop of NITINOL® wire. The 'Sphere' reduces aneurysm inflow by means of a high-density, patterned, elliptical surface that partially occludes the aneurysm neck. The device is secured in the healthy parent vessel by two armatures in the shape of open loops, resulting in negligible disruption of parent or daughter vessel flow. The device is virtually deployed in six anatomically accurate bifurcation aneurysms: three located at the Basilar tip and three located at the terminus bifurcation of the Internal Carotid artery (at the meeting of the middle cerebral and anterior cerebral arteries). Both steady state and transient flow simulations reveal that the device presents with a range of aneurysm inflow reductions, with mean flow reductions falling in the range of 30.6-71.8% across the different geometries. A significant difference is noted between steady state and transient simulations in one geometry, where a zone of flow recirculation is not captured in the steady state simulation. Across all six aneurysms, the device reduces the WSS magnitude within the aneurysm sac, resulting in a hemodynamic environment closer to that of a healthy vessel. We conclude from extensive CFD analysis that the 'Sphere' device offers very significant levels of flow reduction in a number of anatomically accurate aneurysm sizes and locations, with many advantages compared to current clinical cylindrical flow-diverter designs. Analysis of the device's mechanical properties and deployability will follow in future publications.
منابع مشابه
Blood flow in intracranial aneurysms treated with Pipeline embolization devices: computational simulation and verification with Doppler ultrasonography on phantom models
PURPOSE The aim of this study was to validate a computational fluid dynamics (CFD) simulation of flow-diverter treatment through Doppler ultrasonography measurements in patient-specific models of intracranial bifurcation and side-wall aneurysms. METHODS Computational and physical models of patient-specific bifurcation and sidewall aneurysms were constructed from computed tomography angiograph...
متن کاملNew generation of Flow Diverter (surpass) for unruptured intracranial aneurysms: a prospective single-center study in 37 patients.
BACKGROUND AND PURPOSE In patients harboring intracranial aneurysms, the major goal in treatment is to prevent bleeding. A new generation of an endoluminal device (Surpass Flow Diverter [Surpass]) was developed to reconstruct parent artery and occlude the aneurysm. We present our clinical and angiographic single-center experience. METHODS Patients with a wide range of complex unruptured aneur...
متن کاملHemodynamics investigation for a giant aneurysm treated by a flow diverter implantation.
Flow diverter is a kind of stent-like devices with higher metal coverage rate, and it is deployed endovascularly to treat cerebral aneurysms, especially to treat giant aneurysms. However, there exist some arguments about its safety and efficacy. Hemodynamics is considered to play very important role during the initiation, growth and rupture of cerebral aneurysms. In this study, the models of a ...
متن کاملCompacting a Single Flow Diverter versus Overlapping Flow Diverters for Intracranial Aneurysms: A Computational Study.
BACKGROUND AND PURPOSE Locally compacting the mesh of a flow diverter by a dynamic push-pull technique can accelerate intracranial aneurysm healing. We asked how this deployment strategy compares with overlapping 2 flow diverters for aneurysmal flow reduction. MATERIALS AND METHODS Using a high-fidelity virtual stent placement method, we simulated 3 flow-diverter strategies (single noncompact...
متن کاملInitial experience with implantation of novel dual layer flow-diverter device FRED
Flow-diverting stents can help treat complex and wide-necked cerebral aneurysms. The aim of the study was to evaluate initial experiences related to the safety and effectiveness of eight aneurysms treated with a new dual layer coverage designed flow-diverter device. In 2012 Fred flow-diverter devices were used to treat 8 unruptured wide neck (dome-neck ratio ≤ 1.5) and sidewall aneurysms in 6 p...
متن کامل