The effect of limited MR field of view in MR/PET attenuation correction.

نویسندگان

  • Gaspar Delso
  • Axel Martinez-Möller
  • Ralph A Bundschuh
  • Stephan G Nekolla
  • Sibylle I Ziegler
چکیده

PURPOSE A critical question in the development of combined MR/PET scanners is whether MR can provide the tissue attenuation data required for PET reconstruction. Unfortunately, MR images are often unable to encompass the entire patient. The resulting truncation in the transverse plane leads to incomplete attenuation maps, causing artifacts in the reconstructed PET image. This article describes the experiments performed to quantify these artifacts. A method to compensate the missing data was evaluated to determine whether software correction is possible or whether additional transmission hardware has to be included in the scanner. METHODS Three studies were made. First, simulated PET data were used to quantify the bias due to an incomplete attenuation map. A set of spherical lesions was simulated in the lungs and mediastinum of a patient. The data were reconstructed with complete and partial attenuation maps and the uptake differences were evaluated. Second, clinical data from PET/CT oncology patients were used. To reproduce the expected conditions in an MR/PET scanner, only patients scanned with the arms resting along the body were considered. These scans were then used to create maps of the reconstruction bias due to field of view (FOV) limitations. Lastly, a PET reconstruction with incomplete attenuation data was evaluated as a means to obtain attenuation information beyond the MR FOV. The patient outline was automatically segmented with a three-dimensional snake algorithm and used to fill the truncated data in the attenuation map. RESULTS Average bias up to 15% and local biases up to 50% were estimated when PET data were reconstructed with incomplete attenuation information. Completing the attenuation map with data extracted from a PET prereconstruction globally reduced these biases to below 10%. This correction proved to be tolerant to inaccuracies in positioning and attenuation values. However, local artifacts up to 20% could still be found near the edges of the MR FOV. CONCLUSIONS MR FOV restrictions can indeed make the reconstructed PET data unacceptable for diagnostic purposes. Biases can be globally compensated by automatic preprocessing of the attenuation map. However, inaccuracies in the correction will result in small artifacts near the periphery of the image that could lead to false-positive findings.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of state-of-the-art atlas-based bone segmentation approaches from brain MR images for MR-only radiation planning and PET/MR attenuation correction

Introduction: Magnetic Resonance (MR) imaging has emerged as a valuable tool in radiation treatment (RT) planning as well as Positron Emission Tomography (PET) imaging owing to its superior soft-tissue contrast. Due to the fact that there is no direct transformation from voxel intensity in MR images into electron density, itchr('39')s crucial to generate a pseudo-CT (Computed Tomography) image ...

متن کامل

Improvement of attenuation correction in time-of-flight PET/MR imaging with a positron-emitting source.

UNLABELLED Quantitative PET imaging relies on accurate attenuation correction. Recently, there has been growing interest in combining state-of-the-art PET systems with MR imaging in a sequential or fully integrated setup. As CT becomes unavailable for these systems, an alternative approach to the CT-based reconstruction of attenuation coefficients (μ values) at 511 keV must be found. Deriving μ...

متن کامل

Attenuation Effects of MR Headphones During Brain PET/MR Studies.

UNLABELLED PET/MR offers potential advantages over PET/CT that are currently under investigation. One of the challenges of PET/MR is attenuation correction, as there is no simple correlation between MR signal intensity and the attenuation of 511-keV photons detected in PET. Currently, dedicated MR sequences are used to segment voxels into categories that are then assigned a predetermined attenu...

متن کامل

Principles of PET/MR Imaging.

Hybrid PET/MR systems have rapidly progressed from the prototype stage to systems that are increasingly being used in the clinics. This review provides an overview of developments in hybrid PET/MR systems and summarizes the current state of the art in PET/MR instrumentation, correction techniques, and data analysis. The strong magnetic field requires considerable changes in the manner by which ...

متن کامل

Comparison of threshold-based and watershed-based segmentation for the truncation compensation of PET/MR images

Recently introduced combined PET/MR scanners need to handle the specific problem that a limited MR field of view sometimes truncates arm or body contours, which prevents an accurate calculation of PET attenuation correction maps. Such maps of attenuation coefficients over body structures are required for a quantitatively correct PET image reconstruction. This paper addresses this problem by pre...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 37 6Part1  شماره 

صفحات  -

تاریخ انتشار 2010