Effects of Bileaflet Mechanical Mitral Valve Rotational Orientation on Left Ventricular Flow Conditions

نویسندگان

  • John C Westerdale
  • Ronald Adrian
  • Kyle Squires
  • Hari Chaliki
  • Marek Belohlavek
چکیده

We studied left ventricular flow patterns for a range of rotational orientations of a bileaflet mechanical heart valve (MHV) implanted in the mitral position of an elastic model of a beating left ventricle (LV). The valve was rotated through 3 angular positions (0, 45, and 90 degrees) about the LV long axis. Ultrasound scans of the elastic LV were obtained in four apical 2-dimensional (2D) imaging projections, each with 45 degrees of separation. Particle imaging velocimetry was performed during the diastolic period to quantify the in-plane velocity field obtained by computer tracking of diluted microbubbles in the acquired ultrasound projections. The resulting velocity field, vorticity, and shear stresses were statistically significantly altered by angular positioning of the mechanical valve, although the results did not show any specific trend with the valve angular position and were highly dependent on the orientation of the imaging plane with respect to the valve. We conclude that bileaflet MHV orientation influences hemodynamics of LV filling. However, determination of 'optimal' valve orientation cannot be made without measurement techniques that account for the highly 3-dimensional (3D) intraventricular flow.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A D-Shaped Bileaflet Bioprosthesis which Replicates Physiological Left Ventricular Flow Patterns

Prior studies have shown that in a healthy heart, there exist a large asymmetric vortex structure that aids in establishing a steady flow field in the left ventricle. However, the implantation of existing artificial heart valves at the mitral position is found to have a negative effect on this physiological flow pattern. In light of this, a novel D-shaped bileaflet porcine bioprosthesis (GD val...

متن کامل

Direct imaging of bileaflet mechanical valve behavior in the tricuspid position.

OBJECTIVE The optimal orientation of a bileaflet mechanical valve for tricuspid valve replacement (TVR) has not yet been determined. The aim of this study was to use fiberoptic cardioscopy to evaluate the effect of orientation of a mechanical valve implanted in the tricuspid position on bileaflet mechanical valve behavior. METHODS Twelve pigs (50-59 kg) underwent TVR with a St. Jude Mechanica...

متن کامل

Numerical Modeling of Intraventricular Flow during Diastole after Implantation of BMHV

This work presents a numerical simulation of intraventricular flow after the implantation of a bileaflet mechanical heart valve at the mitral position. The left ventricle was simplified conceptually as a truncated prolate spheroid and its motion was prescribed based on that of a healthy subject. The rigid leaflet rotation was driven by the transmitral flow and hence the leaflet dynamics were so...

متن کامل

Independently from mitral regurgitation, Barlow disease may cause left ventricular enlargement and thereby anticipation of surgery.

Sir: We read the article ‘Anatomic characteristics of bileaflet mitral valve prolapse – Barlow disease – in patients undergoing mitral valve repair’ by Rostagno et al. (2014) with great interest. In this article, they shared their surgical observations about bileaflet mitral valve prolapse (MVP) in terms of mitral valve repair. We appreciated their experiences on mitral valve repair in patients...