Central Corneal Thickness Measurement by Ultrasound versus Orbscan II
نویسندگان
چکیده
PURPOSE To compare Orbscan II and ultrasonic pachymetry for measurement of central corneal thickness (CCT) in eyes scheduled for keratorefractive surgery. METHODS CCT was measured using Orbscan II (Bausch & Lomb, USA) and then by ultrasonic pachymetry (Tomey SP-3000, Tomey Ltd, Japan) in 100 eyes of 100 patients with no history of ocular surgery scheduled for excimer laser refractive surgery. RESULTS Mean CCT was 544.7±35.5 (range 453-637) μm by ultrasonic pachymetry versus 546.9±41.6 (range 435-648) μm measured by Orbscan II applying an acoustic factor of 0.92 (P=0.14). The standard deviation of measurements was greater with Orbscan pachymetry but the difference was not statistically significant. CONCLUSION CCT measurements by Orbscan II (applying an acoustic factor) and by ultrasonic pachymetry are not significantly different; however, when CCT readings by Orbscan II are in the lower range, it is advisable to recheck the measurements using ultrasonic pachymetry.
منابع مشابه
Measurement of Central Corneal Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes
PURPOSE To compare ultrasound pachymetry and Orbscan II for measurement of central corneal thickness (CCT) in normal eyes. METHODS The current study was performed at Labbafinejad Medical Center (LMC), Tehran, Iran. Three hundred eyes from 150 healthy individuals referred for keratorefractive surgery were assessed first by Orbscan II and then by ultrasound pachymetry, and CCT values were recor...
متن کاملCentral Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-slit Topography Exclusively in Thin Non-keratoconic Corneas
PURPOSE To evaluate the agreement among Pentacam, Orbscan and ultrasound (US) pachymetry for measurement of central corneal thickness (CCT) in thin corneas with normal topographic pattern. METHODS We included 88 eyes of 44 refractive surgery candidates with thinnest pachymetric readings of 500 micrometers (μm) or less on Orbscan, a normal topographic pattern, no sign of keratoconus, and best ...
متن کاملCentral Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy
PURPOSE To determine inter-device agreement for central corneal thickness (CCT) measurement among ultrasound pachymetry, rotating Scheimpflug imaging (Pentacam, Oculus, Wetzlar, Germany), and scanning slit corneal topography (Orbscan II, Bausch & Lomb, Rochester, NY, USA) in highly myopic eyes before and after photorefractive keratectomy (PRK). METHODS This prospective comparative study inclu...
متن کاملCentral and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry.
PURPOSE To compare thickness measurements of the central 6.0 mm of the cornea obtained with the Orbscan(R) II topography system and topographical ultrasound pachymetry. SETTING School of Optometry, University of Santiago de Compostela, Galicia, Spain. METHODS In 24 right eyes, pachymetric measurements were taken at the center and 1.2 mm and 3.0 mm on the superior and inferior hemimeridians....
متن کاملComparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter
Purpose: To compare thickness of central cornea measured using Pentacam, Orbscan II, and ultrasound pachymeter Methods: Patients with no history of corneal diseases or systemic diseases affecting eyes, who did not wear contact lens or use eye medications, and who with no previous history of corneal surgery were selected for this study. Central corneal thickness (CCT) was measured by three metho...