Performance improvements in a MF-EIT system for acute stroke: The UCL Mk2.5
نویسندگان
چکیده
A recently developed EITS system, the UCLH mk2.5 was tested on a resistor phantom and in linear time and frequency difference imaging of a cylindrical tank, 10cm diameter, filled with 0.1% saline, and 16 electrodes in a ring; the test object was a cylinder of banana, 2 cm long and 1 cm in diameter. Reproducible images with the same location as the test object were obtained with both methods. Noise over time was similar to previous measurements, 0.3%. The pixels located at the centre of the perturbation showed a similar spectrum to that measured with a HP impedance analyser with a centre frequency of 16 kHz compared to the 10 kHz with the HP impedance analyser. Overall, the system is much improved and is being evaluated in blood filled tanks, to represent more demanding multi-frequency test objects, with a view to use in human subjects. Keywords— multi-frequency electrical impedance tomography, EITS, acute stroke detection, calibration.
منابع مشابه
A comparison of two EIT systems suitable for imaging impedance changes in epilepsy.
Electrical impedance tomography (EIT) has the potential to produce functional images of the conductivity changes associated with epilepsy to help localization of epileptic foci. Scalp voltage changes associated with internal conductivity changes due to focal seizures have been shown at the limit of detectability for present EIT systems. The performances of two EIT systems, which may be employed...
متن کاملEffects of Visual Biofeedback Therapy on Postural Balance of Stroke Patients
Purpose: Postural balance deficit is one of the common post-stroke disabilities. Providing visual biofeedback while balance activities are performed is a way to improve postural balance disorders following stroke. But among the research published, there is incoherency about the positive effects of visual biofeedback therapy. The purpose of this study was to investigate the effects of usin...
متن کاملThe effect of constant magnetic field on the formalin-induced pain in mice
Magnetic field (MF) has had many therapeutic usages in traditional medicine. Today, it is used to accelerate the recovery period after surgical operations by some medical centers and in some horse-training farms in the world. There are some scientific reports on the alternative effect of MF on nervous system function, convulsion, pain, neurons, embryonic cells and immune system cells. For this ...
متن کاملThe effect of constant magnetic field on the formalin-induced pain in mice
Magnetic field (MF) has had many therapeutic usages in traditional medicine. Today, it is used to accelerate the recovery period after surgical operations by some medical centers and in some horse-training farms in the world. There are some scientific reports on the alternative effect of MF on nervous system function, convulsion, pain, neurons, embryonic cells and immune system cells. For this ...
متن کاملDesign and calibration of a compact multi-frequency EIT system for acute stroke imaging.
A new, compact UCLH Mk 2.5 EIT system has been developed and calibrated for EIT imaging of the head. Improvements include increased input and output impedances, increased bandwidth and improved CMRR (80 dB) and linearity over frequencies and load (0.2% on a single channel, +/-0.7% on a saline tank over 20 Hz-256 kHz and 10-65 Omega). The accuracy of the system is sufficient to image severe acut...