[Estimation of the phantom scatter factor (Sp) of rectangular fields].
نویسندگان
چکیده
This study proposes a method to accurately estimate the phantom scatter factor (Sp) of arbitrary rectangular fields. We measured output doses in water and air; these measured values were based on square fields and a limited number of symmetric rectangular fields using 4 MV and 10 MV X-rays of a Varian Clinac-iX. We calculated Sp from these measured values. Then, using these Sp values, we estimated equations of Sp on square fields consisting of the primary dose, Day's scatter, and forward scatter. This equation may be used to estimate the Sp value on a square field, but it cannot estimate the Sp value on a rectangular field. We investigated the calculation method for an equivalent square of a rectangular field. As a result, this study's calculation method for an equivalent square, the area ratio correction method, was more accurate than the conventional Bjärngard's method. Therefore, when using the approximate equation of Sp on a square field and the equivalent square calculated by the area ratio correction method, a Sp value of an arbitrary rectangular field may be accurately estimated.
منابع مشابه
Calculation the phantom scatter factor for the linear accelerators device
Introduction: In the last few decades, a lot of monte carlo codes have been introduced for medical applications. this is important because through simulation, we are fully independent of the spectral scattering factor that we are not able measure in real terms. The effect of the radiation energy, the dimension of the radiation field, the sensitive volume of the ion chamber on t...
متن کاملA new approach to scatter correction in SPECT images based on Klein_Nishina equation
Introduction: Scattered photon is one of the main defects that degrade the quality and quantitative accuracy of nuclear medicine images. Accurate estimation of scatter in projection data of SPECT is computationally extremely demanding for activity distribution in uniform and non-uniform dense media. Methods: The objective of this paper is to develop and validate a scatter correction technique ...
متن کاملAn Analytical Method to Calculate Phantom Scatter Factor for Photon Beam Accelerators
INTRODUCTION One of the important input factors in the commissioning of the radiotherapy treatment planning systems is the phantom scatter factor (Sp) which requires the same collimator opening for all radiation fields. In this study, we have proposed an analytical method to overcome this issue. METHODS The measurements were performed using Siemens Primus Plus with photon energy 6 MV for fiel...
متن کاملOut-Of-Field Dose Measurement and Second Cancer-Risk Estimation Following External Beam Radiotherapy and Brachytherapy for Cervical Cancer Treatment: A Phantom Study
Introduction: The present study aimed to measure the scatter and leakage dose received by out-of-field organs while delivering Radiotherapy (RT) treatment of cervical cancer. Moreover, this study estimated the risk of second cancer (SC). The doses to out-of-field organs were measured using a lithium fluoride (TLD 100) dosimeter while delivering External Beam Radiothera...
متن کاملMeasurement and comparison of head scatter factor for 9MV photon beam using the build-up cap and a columnar mini-phantom
Introduction: The Sc is an important parameter to monitor unit (MU) calculation in radiation therapy procedure. The aim of this study is to measurement and comparison of head scatter factor (Sc) for 9MV photon beam using the build-up cap and a columnar mini-phantom. Materials and Methods: The measurements performed using a calibrated ion chamber (Scanditron...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Nihon Hoshasen Gijutsu Gakkai zasshi
دوره 68 1 شماره
صفحات -
تاریخ انتشار 2012