Trueness Assessment for Serum Glucose Measurement Using Commercial Systems through the Preparation of Commutable Reference Materials

نویسندگان

  • ChangYu Xia
  • Ou Liu
  • LanZhen Wang
  • GuoBing Xu
چکیده

BACKGROUND Commutable reference materials (RMs) are suitable for end-users for evaluating the metrological traceability of values obtained using routine measurement systems. We assessed the performance of 6 routine measurement systems with validated secondary RMs. METHODS We tested the homogeneity, stability, and commutability of 5 minimally processed human serum pools according to the standard guidelines. The serum pools were assigned values as per the reference procedure of the United States Centers for Disease Control and were used to evaluate the trueness of results from 6 commercial measurement systems based on enzymatic methods: 3 glucose oxidase (GOD) and 3 hexokinase (HK) methods. RESULTS The prepared RMs were validated to be sufficiently homogenous, stable, and commutable with the patient samples. Method bias varied for different systems: GOD01, -0.17 to 2.88%; GOD02, 1.66 to 4.58%; GOD03, -0.17 to 3.14%; HK01, -3.48 to -0.85%; HK02, -3.83 to -0.11%, and HK03, -1.82 to -0.27%. CONCLUSIONS We observed that the prepared serum glucose RMs were qualified for trueness assessment. Most of the measurement systems met the minimal quality specifications.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of the Trueness and Inter-Laboratory Precision of Routine Uric Acid Assays Using 4 Frozen Pooled Serum Samples Measured by the Japan Society of Clinical Chemistry's HPLC Method

BACKGROUND Reference procedures are required for evaluating the accuracy of routine analytical systems for uric acid (UA). External quality assessment (EQA) for UA has only been conducted with quality controls in China, and the results have not been published. This study was designed to investigate both the trueness and inter-laboratory precision of UA measurements among routine analytical syst...

متن کامل

Trueness verification and traceability assessment of results from commercial systems for measurement of six enzyme activities in serum: an international study in the EC4 framework of the Calibration 2000 project.

BACKGROUND The in vitro diagnostics directive of the European Union requires traceability to higher order reference measurement procedures and materials for analytes in assuring the result trueness and comparability of laboratory measurements. Manufacturers must ensure that the systems they market are calibrated against available reference systems. Validation of metrologically traceable calibra...

متن کامل

Commutable calibrator with value assigned by the IFCC reference procedure to harmonize serum lactate dehydrogenase activity results measured by 2 different methods.

BACKGROUND The availability of commutable calibrator materials may ease considerably the task of harmonizing assay results and ensuring their traceability to reference procedures. We sought to verify the commutability of potential calibrator materials and evaluate their effectiveness in harmonizing LDH results by 2 measurement methods. METHODS We measured LDH in 109 serum samples and 31 mater...

متن کامل

Trueness verification of actual creatinine assays in the European market demonstrates a disappointing variability that needs substantial improvement. An international study in the framework of the EC4 creatinine standardization working group.

BACKGROUND The European In Vitro Diagnostics (IVD) directive requires traceability to reference methods and materials of analytes. It is a task of the profession to verify the trueness of results and IVD compatibility. METHODS The results of a trueness verification study by the European Communities Confederation of Clinical Chemistry (EC4) working group on creatinine standardization are descr...

متن کامل

State of the art in trueness and interlaboratory harmonization for 10 analytes in general clinical chemistry.

CONTEXT Harmonization and standardization of results among different clinical laboratories is necessary for clinical practice guidelines to be established. OBJECTIVE To evaluate the state of the art in measuring 10 routine chemistry analytes. DESIGN A specimen prepared as off-the-clot pooled sera and 4 conventionally prepared specimens were sent to participants in the College of American Pa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 32  شماره 

صفحات  -

تاریخ انتشار 2012