Comparison of multi‐institutional Varian ProBeam pencil beam scanning proton beam commissioning data
نویسندگان
چکیده
PURPOSE Commissioning beam data for proton spot scanning beams are compared for the first two Varian ProBeam sites in the United States, at the Maryland Proton Treatment Center (MPTC) and Scripps Proton Therapy Center (SPTC). In addition, the extent to which beams can be matched between gantry rooms at MPTC is investigated. METHOD Beam data for the two sites were acquired with independent dosimetry systems and compared. Integrated depth dose curves (IDDs) were acquired with Bragg peak ion chambers in a 3D water tank for pencil beams at both sites. Spot profiles were acquired at different distances from the isocenter at a gantry angle of 0° as well as a function of gantry angles. Absolute dose calibration was compared between SPTC and the gantries at MPTC. Dosimetric verification of test plans, output as a function of gantry angle, monitor unit (MU) linearity, end effects, dose rate dependence, and plan reproducibility were compared for different gantries at MPTC. RESULTS The IDDs for the two sites were similar, except in the plateau region, where the SPTC data were on average 4.5% higher for lower energies. This increase in the plateau region decreased as energy increased, with no marked difference for energies higher than 180 MeV. Range in water coincided for all energies within 0.5 mm. The sigmas of the spot profiles in air were within 10% agreement at isocenter. This difference increased as detector distance from the isocenter increased. Absolute doses for the gantries measured at both sites were within 1% agreement. Test plans, output as function of gantry angle, MU linearity, end effects, dose rate dependence, and plan reproducibility were all within tolerances given by TG142. CONCLUSION Beam data for the two sites and between different gantry rooms were well matched.
منابع مشابه
Commissioning of the world's first compact pencil‐beam scanning proton therapy system
This paper summarizes clinical commissioning of the world's first commercial, clinically utilized installation of a compact, image-guided, pencil-beam scanning, intensity-modulated proton therapy system, the IBA Proteus® ONE, at the Willis-Knighton Cancer Center (WKCC) in Shreveport, LA. The Proteus® ONE is a single-room, compact-gantry system employing a cyclotron-generated proton beam with im...
متن کاملEvaluation of detectors for acquisition of pristine depth‐dose curves in pencil beam scanning
Acquisition of quasi-monoenergetic ("pristine") depth-dose curves is an essential task in the frame of commissioning and quality assurance of a proton therapy treatment head. For pencil beam scanning delivery modes this is often accomplished by measuring the integral ionization in a plane perpendicular to the axis of an unscanned beam. We focus on the evaluation of three integral detectors: two...
متن کاملA Review of Proton Radiation Therapy and the Path to Widespread Clinical Adoption
After many years of development, proton therapy is finally reaching the point of mass adoption in clinical practice. Advances in particle accelerator technology and improved dose delivery techniques have provided strong driving forces for expanded use. Pencil beam scanning (PBS) is the generic name for radiation dose delivery to a target volume using individually controlled small pencil beams o...
متن کاملEffects of Defining Realistic Compositions of the Ocular Melanoma on Proton Therapy
Background: Recent studies in eye plaque brachytherapy have shown a considerable difference between the dosimetric results using water phantom and a model of human eye containing realistic materials. In spite of this fact, there is a lack of simulation studies based on such a model in proton therapy literatures. In the presented work, the effect of utilizing an eye model with ocular media on pr...
متن کاملIba Proton Pencil Beam Scanning: an Innovative Solution for Cancer Treatment
Particles are used in medical application for cancer treatment. Radiotherapy aims to deliver a high dose in a well-defined target volume while sparing the healthy surrounding tissues. Compared to the conventional radiotherapy (X-ray), protons show superior clinical performances thanks to their inherent depth dose characteristic (Bragg peak). In proton therapy, several techniques can be used to ...
متن کامل