Evaluation of imprecision for cardiac troponin assays at low-range concentrations.
نویسندگان
چکیده
BACKGROUND The European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction (MI) has recommended that an increased cardiac troponin should be defined as a measurement above the 99th percentile value of the reference group. A total imprecision (CV) at the decision limit of </=10% is recommended. However, peer-reviewed data on assay imprecision are lacking. The purpose of this study was to construct the clinically relevant imprecision profiles for each of the commercially available cardiac troponin assays. Pools of human sera containing cardiac troponin concentrations around the MI decision limit were assessed to identify the lowest concentration associated with a 10% CV. METHODS Eight serum pools targeting different concentrations of cardiac troponin (I and T) were prepared and stored at -70 degrees C until usage. The cardiac troponin measurement protocol consisted of two replicates per specimen per run, and one run per day for 20 days, using two reagent lots and three calibrations. Manufacturers of each cardiac troponin assay directly performed the measurements. Data analysis for each assay was centralized and performed according to the NCCLS EP5-A guideline. RESULTS The lowest concentrations (microg/L) providing a 10% CV were as follows: AxSYM, 1.22; ACS:180, 0.37; Centaur, 0.33; Immuno 1, 0.34; Access, 0.06; Vidas, 0.36; Liaison, 0.065; Dimension, 0.26; Opus, 0.90; Stratus CS, 0.10; Immulite, 0.32; Vitros ECi, 0.44; Elecsys, 0.04; AIA 21, 0.09. CONCLUSION No cardiac troponin assay was able to achieve the 10% CV recommendation at the 99th percentile reference limit defined by the manufacturer.
منابع مشابه
99th percentile and analytical imprecision of troponin and creatine kinase-MB mass assays: an objective platform for comparison of assay performance.
Cardiac troponins in blood are the most sensitive and specific biochemical markers of myocardial damage and are paramount for classification, risk stratification, and customized therapy in patients with acute coronary syndromes (1, 2). Despite the overt advantages, some important obstacles to troponin analysis and interpretation have remained, such as assay standardization, interference, preana...
متن کاملEdging closer to early optimal patient management with high-sensitivity cardiac troponin assay.
These are exciting times in the evolution of cardiac biomarkers. In the past 12 years, we have witnessed the transition from creatine kinase-MB to cardiac troponin I (cTnI) and T (cTnT) as the standard biomarkers to aid in the diagnosis of acute myocardial infarction (MI). This is evidenced by the overwhelming endorsement through recommendations from laboratory medicine,1 cardiology,2 and emerg...
متن کاملClinical and analytical performance of the liaison cardiac troponin I assay in unstable coronary artery disease, and the impact of age on the definition of reference limits. A FRISC-II substudy.
BACKGROUND Measurements of cardiac troponins are currently used as the standard for the detection of myocardial injury. None of the current assays complies with the new requirements on assay imprecision as proposed by the European Society of Cardiology/American College of Cardiology. Our aim was to evaluate the clinical and analytical performance of the Liaison cardiac troponin I (cTnI) assay. ...
متن کاملHigh-sensitivity cardiac troponin I assays in the diagnosis of acute myocardial infarction
Chest pain is one of the most common presenting symptoms in patients attending the emergency department; yet, less than one in five patients receive a final diagnosis of acute myocardial infarction. In patients without ST segment elevation, early exclusion of acute coronary syndrome has the potential to reduce healthcare expenditure in a population with significant resource implications. Previo...
متن کاملBeing rational about (im)precision: a statement from the Biochemistry Subcommittee of the Joint European Society of Cardiology/American College of Cardiology Foundation/American Heart Association/World Heart Federation Task Force for the definition of myocardial infarction.
We have for many years advocated for very precise cardiac troponin assays to improve the sensitivity of detection of cardiac necrosis and to decrease over time the amount of change (or delta) in cardiac troponin concentrations needed to be considered significant (1, 2 ). During the time period when assays were insufficiently precise at very low levels, we and others advocated using as a cutoff ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 50 2 شماره
صفحات -
تاریخ انتشار 2004