Influence of overnight orthokeratology lens fitting decentration on corneal topography reshaping
نویسندگان
چکیده
Background This retrospective study was designed to investigate the sole influence of orthokeratology (OK) lens fitting decentration on the Zernike coefficients of the reshaped anterior corneal surface. Methods This study comprised a review of 106 right eyes and measurements of corneal topography both before OK and at 1-month follow-up visit. A routine was designed to calculate local corneal surface astigmatism and assist the determination of OK lens fitting decentration from pupil center. The pupil-centered corneal Zernike coefficients of baseline (PCCB) and post-treatment (PCCP) were calculated. Meanwhile, the OK-lens-centered corneal Zernike coefficients of post-treatment (OCCP) were also calculated and considered as the presumptive ideal fitting group without decentration. Relationships between lens fitting decentration and the change of Zernike coefficients including (PCCP - PCCB) and (PCCP - OCCP) were analyzed. Results Patients with a mean age of 11 ± 2.36 years old had an average spherical equivalent refractive error of -3.52 ± 1.06 D before OK. One month after treatment, OK lens fitting decentration from pupil center was 0.68 ± 0.35 mm. RMS of 3rd-order (P < 0.05), RMS of 4th-order (P < 0.001) and RMS of total high order (P < 0.001) corneal Zernike coefficients were increased in PCCP by comparing with OCCP, which was solely caused by lens fitting decentration. Nevertheless, no significant difference was observed in [Formula: see text] (P > 0.05). For the high order corneal Zernike coefficients in (PCCP - OCCP), radial distance of decentration was correlated with [Formula: see text] (r = -0.296, P < 0.05), [Formula: see text] (r = -0.396, P < 0.001), and [Formula: see text] (r = 0.449, P < 0.001), horizontal decentration was significantly correlated with [Formula: see text] (r = 0.901, P < 0.001) and [Formula: see text] (r = 0.340, P < 0.001), and vertical decentration was significantly correlated with [Formula: see text] (r = 0.904, P < 0.001). Conclusions OK lens fitting decentration within 1.5 mm hardly influenced the change of corneal spherical power for myopia correction, but significantly induced additional corneal high order Zernike coefficients including [Formula: see text], [Formula: see text], [Formula: see text], and [Formula: see text].
منابع مشابه
Corneal changes and wavefront analysis after orthokeratology fitting test.
PURPOSE To evaluate corneal changes and ocular aberrations during an orthokeratology test. DESIGN A prospective, nonrandomized cohort study. METHODS Fourteen myopic patients (26 eyes) underwent an orthokeratology fitting test with the BE contact lens (Ultravision Pty, Ltd, Brisbane, Australia). Best spectacle-corrected visual acuity (BSCVA), uncorrected (Ultravision Pty, Ltd, Brisbane, Aust...
متن کاملVertical and horizontal thickness profiles of the corneal epithelium and Bowman's layer after orthokeratology.
PURPOSE To investigate thickness profile changes of the corneal epithelium and Bowman's layer at the vertical and horizontal meridians with overnight myopia orthokeratology (OK) lenses. METHODS Twenty subjects (age range: 19-33 years) wore reverse-geometry rigid gas-permeable OK lenses in both eyes for 30 days. Before lens wear and after 1, 7, and 30 days of overnight lens wear, evaluation of...
متن کاملPredictive role of corneal Q-value differences between nasal–temporal and superior–inferior quadrants in orthokeratology lens decentration
BACKGROUND To investigate the association between pretreatment corneal parameters and orthokeratology lens decentration. METHODS This retrospective study included a total of 108 eyes in 60 myopia patients, who were divided into a lens-decentration and a control group. Various pretreatment corneal parameters were analyzed by receiver operating characteristic curves (ROC curves), including corn...
متن کاملThe Children's Overnight Orthokeratology Investigation (COOKI) pilot study.
PURPOSE Innovations in contact lens materials and designs allow patients to wear contact lenses during sleep to flatten the cornea and temporarily to reduce myopic refractive error and improve unaided visual acuity. We conducted the Children's Overnight Orthokeratology Investigation (COOKI) pilot study, a case series, to describe the refractive error and visual changes, as well as the slitlamp ...
متن کاملCorneal thickness changes in hyperopic orthokeratology measured by optical pachometry.
PURPOSE To investigate the time course of corneal thickness changes in overnight hyperopic orthokeratology (OK) lens wear for a 4-day lens-wearing period. METHODS Fourteen subjects (age range, 20-37 years) were fitted with hyperopic OK lenses in one eye only. The fellow eye acted as a non-lens wearing control. Lenses were worn overnight only for 4 nights, and changes from baseline in subjecti...
متن کامل