Safety analysis of proposed data-driven physiologic alarm parameters for hospitalized children.
نویسندگان
چکیده
INTRODUCTION Modification of alarm limits is one approach to mitigating alarm fatigue. We aimed to create and validate heart rate (HR) and respiratory rate (RR) percentiles for hospitalized children, and analyze the safety of replacing current vital sign reference ranges with proposed data-driven, age-stratified 5th and 95th percentile values. METHODS In this retrospective cross-sectional study, nurse-charted HR and RR data from a training set of 7202 hospitalized children were used to develop percentile tables. We compared 5th and 95th percentile values with currently accepted reference ranges in a validation set of 2287 patients. We analyzed 148 rapid response team (RRT) and cardiorespiratory arrest (CRA) events over a 12-month period, using HR and RR values in the 12 hours prior to the event, to determine the proportion of patients with out-of-range vitals based upon reference versus data-driven limits. RESULTS There were 24,045 (55.6%) fewer out-of-range measurements using data-driven vital sign limits. Overall, 144/148 RRT and CRA patients had out-of-range HR or RR values preceding the event using current limits, and 138/148 were abnormal using data-driven limits. Chart review of RRT and CRA patients with abnormal HR and RR per current limits considered normal by data-driven limits revealed that clinical status change was identified by other vital sign abnormalities or clinical context. CONCLUSIONS A large proportion of vital signs in hospitalized children are outside presently used norms. Safety evaluation of data-driven limits suggests they are as safe as those currently used. Implementation of these parameters in physiologic monitors may mitigate alarm fatigue. Journal of Hospital Medicine 2015;11:817-823. © 2015 Society of Hospital Medicine.
منابع مشابه
Implementation of Data Drive Heart Rate and Respiratory Rate parameters on a Pediatric Acute Care Unit
The majority of hospital physiologic monitor alarms are not clinically actionable and contribute to alarm fatigue. In 2014, The Joint Commission declared alarm safety as a National Patient Safety Goal and urged prompt action by hospitals to mitigate the issue [1]. It has been demonstrated that vital signs in hospitalized children are quite different from currently accepted reference ranges [2]....
متن کاملRole of Large Clinical Datasets From Physiologic Monitors in Improving the Safety of Clinical Alarm Systems and Methodological Considerations: A Case From Philips Monitors
BACKGROUND Large datasets of the audit log of modern physiologic monitoring devices have rarely been used for predictive modeling, capturing unsafe practices, or guiding initiatives on alarm systems safety. OBJECTIVE This paper (1) describes a large clinical dataset using the audit log of the physiologic monitors, (2) discusses benefits and challenges of using the audit log in identifying the...
متن کاملVideo methods for evaluating physiologic monitor alarms and alarm responses.
False physiologic monitor alarms are extremely common in the hospital environment. High false alarm rates have the potential to lead to alarm fatigue, leading nurses to delay their responses to alarms, ignore alarms, or disable them entirely. Recent evidence from the U.S. Food and Drug Administration (FDA) and The Joint Commission has demonstrated a link between alarm fatigue and patient deaths...
متن کاملImpact of a Safety Huddle-Based Intervention on Monitor Alarm Rates in Low-Acuity Pediatric Intensive Care Unit Patients.
BACKGROUND Physiologic monitors generate high rates of alarms in the pediatric intensive care unit (PICU), yet few are actionable. OBJECTIVE To determine the association between a huddle-based intervention focused on reducing unnecessary alarms and the change in individual patients' alarm rates in the 24 hours after huddles. DESIGN Quasi-experimental study with concurrent and historical con...
متن کاملData-driven Adaptive Safety Monitoring Using Virtual Subjects in Medical Cyber-Physical Systems: A Glucose Control Case Study
Medical cyber-physical systems (MCPS) integrate sensors, actuators, and software to improve patient safety and quality of healthcare. These systems introduce major challenges to safety analysis because the patient’s physiology is complex, nonlinear, unobservable, and uncertain. To cope with the challenge that unidentified physiological parameters may exhibit short-term variances in certain clin...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of hospital medicine
دوره 11 12 شماره
صفحات -
تاریخ انتشار 2016