Determining the critical effective temperature and heat dispersal pattern in monopolar radiofrequency ablation using temperature-time integration.
نویسندگان
چکیده
The radiofrequency ablation (RFA) lesion size is posited to be disproportionate to the total delivered energy, and temperature-time integration (TTI) may have a more critical effect on lesion size. The present study aimed to evaluate this hypothesis by determining the temperature threshold and temperature distribution over tissues during the RFA lesioning process. Using an ex vivo chicken tissue model and an in vivo rabbit model with RFA applied for 2 min under various target temperature settings, the resultant lesions were evaluated histologically using Masson's trichrome stain. The temperature distribution over the tissue during the RFA lesioning process was also determined using a VT02 Visual IR Thermometer. It was revealed that the thermal injury threshold for RFA in the chicken tissues was ~65°C, but that it ranged from 55-65°C in mammals. Using infra-red thermal imaging, the temperature gradient (from the center to the periphery) during the RFA lesioning process demonstrated a uniform heat diffusion pattern. This data supports the proposed hypothesis that TTI is a critical parameter in determining RFA lesion size and can be applied clinically using the following equation: [Target temperature - 55 (°C)] × time (sec) is proportional to RFA lesion size.
منابع مشابه
Heat Sink Effect on Tumor Ablation Characteristics as Observed in Monopolar Radiofrequency, Bipolar Radiofrequency, and Microwave, Using Ex Vivo Calf Liver Model
Thermal ablation of liver tumors near large blood vessels is affected by the cooling effect of blood flow, leading to incomplete ablation. Hence, we conducted a comparative investigation of heat sink effect in monopolar (MP) and bipolar (BP) radiofrequency ablation (RFA), and microwave (MW) ablation devices.With a perfused calf liver, the ablative performances (volume, mass, density, dimensions...
متن کاملAn Investigation of Brightness Changes in Ultrasound Images due to Temperature Variations in Liver Tissue during Radiofrequency Ablation
Introduction: one of the minimally invasive methods of treatmenting liver malignancies is Radiofrequency Ablation (RFA) which can be applied to primary or secondary tumors. Nevertheless, the disease recurrence is probable after RFA application partially due to the inadequate capability of temperature monitoring of the target tissue and visualizing the thermal damage. The goal of this work was t...
متن کاملRelative brightness due to temperature changes with ultrasound image analysis during Radiofrequency ablation
Introduction: Diagnosis of primary and secondary cancers to treat patients with diffuse and malignant tumors is an important subject. The appropriate treatment will be eliminating primary and metastatic tumors. Radiofrequency ablation with localized heat induction in the target tissue causes irreversible cellular damage and tissue coagulation. Estimation of tissue temperature c...
متن کاملبررسی اثر رادیوفرکوئنسی لوزه ها در درمان بوی بد دهان ناشی از التهاب مزمن لوزه همراه با کازئوم
Introduction & Objective: Chronic tonsillitis with caseum has been proven to be the cause of halitosis. The objective of this study was to evaluate the efficacy and safety of temperature-controlled radiofrequency tonsil ablation in the treatment of halitosis caused by chronic tonsil-litis with caseum. Materials & Methods: In this clinical trial study 28 adult patients with halitosis due to ch...
متن کاملComparison of the presence and non-presence states of magnetite nanoparticles in tissue-equivalent breast phantom via radiofrequency hyperthermia
Objective(s): Breast cancer is a fatal disease and the leading cause of mortality in women. Radiofrequency hyperthermia is an approach to the treatment of cancer cells through increasing their temperature. The present study aimed to investigate breast tumor ablation via radiofrequency hyperthermia in the presence and non-presence states of magnetite nanoparticles and assess the effects of magne...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Experimental and therapeutic medicine
دوره 11 3 شماره
صفحات -
تاریخ انتشار 2016