64-element intraluminal ultrasound cylindrical phased array for transesophageal thermal ablation under fast MR temperature mapping: an ex vivo study.
نویسندگان
چکیده
This work was undertaken to investigate the feasibility of using a cylindrical phased array for transoesophaeal thermal ablation under magnetic resonance (MR) imaging guidance. Sixty-four transducers (0.45 mm wide by 15 mm tall), operating at 4.6 MHz, were spread around the periphery of a 10.6-mm-diam cylinder. The head of the applicator was covered with a 65-microm thick latex balloon attached using watertight seals. This envelope was inflated with degassed water to provide acoustic coupling between the transducer and the tissues. The underlying operating principle of this applicator is to rotate a plane ultrasound beam electronically. For this purpose, eight adjacent transducers were excited with appropriate delay times so as to generate a plane wave. The exposure direction was changed by exciting a different set of eight elements. Ex vivo experiments conducted on 47 samples of pig liver under MR temperature monitoring demonstrated the ability of this applicator to generate cylindrical or sector-based coagulation necroses at depths up to 19 mm with excellent angular precision by applying 20 W/cm2. MR thermometry was performed in "real-time" with segmented echo-planar imaging gradient echo sequences. The temporal resolution was approximately 3 s/ image. The average value for the temperature baseline in liver tissue close to the applicator was 0.3 degrees C (+/- 0.6 degrees C). The thermal dose delivered in tissues was computed on-line during temperature imaging. Excellent MR compatibility was demonstrated, all MR acquisitions were performed without susceptibility artifacts or radio-frequency interferences with the ultrasound device. Thermal lesions identified on post-treatment follow up showed good correlation with online MR thermometry data. The individual differences between measurements performed visually and using MRI thermal dose maps were about 11% of volume. This study demonstrated the feasibility of thermal ablation using a phased array intraluminal ultrasound applicator and on-line MR monitoring.
منابع مشابه
500-element ultrasound phased array system for noninvasive focal surgery of the brain: a preliminary rabbit study with ex vivo human skulls.
The aim of this study was to test a prototype MRI-compatible focused ultrasound phased array system for trans-skull brain tissue ablation. Rabbit thigh muscle and brain were sonicated with a prototype, hemispherical 500-element ultrasound phased array operating at frequencies of 700-800 kHz. An ex vivo human skull sample was placed between the array and the animal tissue. The temperature elevat...
متن کاملSession: 1J
Purpose: To test a prototype Magnetic Resonance Imaging (MRI) compatible focused ultrasound phased array system for trans-skull brain tissue ablation using ex vivo human skulls and an animal model. Methods: Rabbit thigh muscle (N=3) and brain (N=10) were sonicated with a prototype, hemispherical 500element ultrasound phased array operating at frequencies between 700 and 800 kHz. An ex vivo huma...
متن کاملEndocavitary phased array applicator of therapeutic ultrasound with an integrated opposed-solenoid coil for high resolution MRI-guided thermotherapy: an in vivo study
Introduction. Minimally invasive treatment using endoluminal high intensity contact ultrasound (HICU) is being suggested as an alternative cancer therapy in esophagus, colon or rectum [1]. This ablative technique uses the acoustic energy delivered by an ultrasound transducer introduced in the lumen, close to the target tissue. A successful therapy requires an active control of the energy deposi...
متن کاملPre-clinical testing of a phased array ultrasound system for MRI-guided noninvasive surgery of the brain--a primate study.
MRI-guided and monitored focused ultrasound thermal surgery of brain through intact skull was tested in three rhesus monkeys. The aim of this study was to determine the amount of skull heating in an animal model with a head shape similar to that of a human. The ultrasound beam was generated by a 512 channel phased array system (Exablate 3000, InSightec, Haifa, Israel) that was integrated within...
متن کاملMulti-element Ultrasound Phased Array Applicator for the Ablation of Deep-seated Tissue
High-intensity focused ultrasound (HIFU) has attracted increasing interests as a promising noninvasive modality for the treatment of deep tumors in the thoracic and abdominal cavity. A 90-element HIFU spherical phased array applicator operated at 1MHz has been developed for deep tissue ablation. The spherical array with a 5 cm wide central hole has a 21 cm diameter and an 18 cm radius of curvat...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 33 8 شماره
صفحات -
تاریخ انتشار 2006