Performance Evaluation of Fast Microfluidic Thermal Lysis of Bacteria for Diagnostic Sample Preparation †
نویسندگان
چکیده
Development of new diagnostic platforms that incorporate lab-on-a-chip technologies for portable assays is driving the need for rapid, simple, low cost methods to prepare samples for downstream processing or detection. An important component of the sample preparation process is cell lysis. In this work, a simple microfluidic thermal lysis device is used to quickly release intracellular nucleic acids and proteins without the need for additional reagents or beads used in traditional chemical or mechanical methods (e.g., chaotropic salts or bead beating). On-chip lysis is demonstrated in a multi-turn serpentine microchannel with external temperature control via an attached resistive heater. Lysis was confirmed for Escherichia coli by fluorescent viability assay, release of ATP measured with bioluminescent assay, release of DNA measured by fluorometry and qPCR, as well as bacterial culture. Results comparable to standard lysis techniques were achievable at temperatures greater than 65 °C and heating durations between 1 and 60 s.
منابع مشابه
Microfluidic blood sample preparation for rapid sepsis diagnostics
Sepsis, commonly referred to as blood poisoning, is a serious medical condition characterized by a whole-body inflammatory state caused by microbial infection. Rapid treatment is crucial, however, traditional culturebased diagnostics usually takes 2-5 days. The overall aim of the thesis is to develop microfluidic based sample preparation strategies, capable of isolating bacteria from whole bloo...
متن کاملElectrophoretic Concentration and Electrical Lysis of Bacteria in a Microfluidic Device Using a Nanoporous Membrane
Pathogenic bacteria such as Escherichia coli O157, Salmonella and Campylobacter are the main causes for food and waterborne illnesses. Lysis of these bacteria is an important component of the sample preparation for molecular identification of these pathogens. The pathogenicity of these bacteria is so high that they cause illness at very low concentrations (1–10 CFU/100 mL). Hence, there is a ne...
متن کاملEvaluation of a nanosuspension formulation prepared through microfluidic reactors for pulmonary delivery of budesonide using nebulizers
This study aimed to determine the aerosolization behavior of a nanodispersion of budesonide, prepared using microfluidic reactors. The size and morphology of budesonide nanoparticles were characterized by photon correlation spectroscopy (PCS) and transmission electron microscopy (TEM). Processing/formulation parameters for formation of the nanoparticles were studied to determine their effects o...
متن کاملEvaluation of a nanosuspension formulation prepared through microfluidic reactors for pulmonary delivery of budesonide using nebulizers
This study aimed to determine the aerosolization behavior of a nanodispersion of budesonide, prepared using microfluidic reactors. The size and morphology of budesonide nanoparticles were characterized by photon correlation spectroscopy (PCS) and transmission electron microscopy (TEM). Processing/formulation parameters for formation of the nanoparticles were studied to determine their effects o...
متن کاملTowards Point-of-care Diagnostics: a Microfluidic Sample Preparation Chip for Concentration of Bacteria and Rna Extraction
We report on the development and proof-of-concept of a microfluidic chip. It provides a full range of pretreatment steps for the point-of-care detection of bacteria with a total process time of less than 15 minutes. The device integrates concentration, lysis and nucleic acid purification. Successful operation is demonstrated using E. coli as a model organism.