Lactulose interferes in the alkaline picrate assay for creatinine.
نویسنده
چکیده
Errors of more than 10 mg/L in measured serum creatinine concentrations were encountered for an azotemic patient who was given lactulose orally. The magnitude of the apparent error fluctuated with the dose of lactulose. Additions of lactulose to serum produced a linear increase in the creatinine measured by each of two automated methods that involve use of the alkaline picrate (Jaffé) reaction. A lactulose concentration of 100 g/L produced positive interferences of 30 and 65 mg/L in kinetic (Beckman Astra) and continuous-flow (Technicon SMAC) assays, respectively, but caused no problem in an enzymatic assay for creatinine. The results of creatinine assays must be interpreted with caution in patients treated with lactulose.
منابع مشابه
Nitromethane interferes in assay of creatinine by the Jaffé reaction.
A patient who ingested a methanol/nitromethane mixture (model airplane fuel) showed an apparent plasma creatinine concentration of 8.0 mmol/L by the Jaffé reaction when the actual creatinine concentration, as measured by a specific enzymatic method, was 0.09 mmol/L. This effect was due to nitromethane in the plasma. Interference by nitromethane with the determination of creatinine by reaction w...
متن کاملKinetic potentiometric determination of creatinine in serum with a picrate ion-selective membrane electrode.
We describe a new kinetic method for potentiometric determination of creatinine in serum, based on the creatinine--picrate reaction in alkaline medium (Jaffé reaction). The reaction is monitored with a picrate-selective electrode, and the increase in electrode potential during 270 s is measured and related directly to the creatinine concentration. Small cation-exchange columns are used to separ...
متن کاملA simplified method for HPLC determination of creatinine in mouse serum.
Mouse models are frequently used to study renal function. However, mouse serum contains chromagens that interfere with standard picric acid-based assays for serum creatinine. Several alternative methods exist for serum creatinine measurements, including assay by high-performance liquid chromatography (HPLC), but only one has been adapted to mouse serum. Creatinine was measured in serum by aceto...
متن کاملThe estimation of creatinine in serum.
Although the clearance of endogenous creatinine as a measure of the glomerular filtration rate has been much criticized, it remains the only possible measure where repeated or continuous investigation is needed, and has already yielded much useful information. When we started to use it, however, it became apparent that while the estimation of creatinine in urine was satisfactory, that of plasma...
متن کاملReaction of picrate with creatinine and cepha antibiotics.
The concentration of creatinine in serum, which is used to estimate glomerular filtration rate, is measured by reaction with alkaline picrate, but this reaction is not specific for creatinine. Although several other cephalosporin antibiotics have been reported not to react with picrate, we reacted picrate with creatinine, cefoxitin, penicillin, and eight different cephalosporins, and found that...