Shuttle dose at the Vienna Leksell Gamma Knife.

نویسندگان

  • A Ertl
  • M Zehetmayer
  • A Schöggl
  • K Kitz
  • R Koschuch
  • H Stadtmann
  • P Kindl
چکیده

The aim of this study was to determine the shuttle dose for all collimator helmets (4, 8, 14 and 18 mm) of the Gamma Knife, model B, in Vienna, Austria. The additional dose accumulated during the transport of the patient in and out of the treatment position should be considered in the dose planning procedure of multicentre treatment regimens and in fractionated stereotactic Gamma Knife radiotherapy. The GafChromic film study was basically used to determine the shuttle dose of all four collimator helmets. In addition, measurements with an ionization chamber (18 and 14 mm collimator--and, for the 18 mm collimator helmet, TLD dosimetry--were performed in order to confirm the GafChromic film data. The shuttle dose ranged between 99.6 and 183.5 mGy, depending mainly on the size of the collimator and the irradiated isocentres at the half-life activity of Co-60 in a brand new Gamma unit. Our film-generated data were in good correlation with the dose levels obtained with the ionization chamber and the TLD dosimetry, showing a dose difference of less than 0.8%. Since it was possible to verify the shuttle dose even for the 4 and 8 mm collimator helmets, we consider it a non-negligible factor and would advocate the inclusion of the shuttle dose in radiosurgical dose planning.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of basic physical and dosimetric parameters of synthetic single-crystal diamond detector and its use in Leksell Gamma Knife and CyberKnife small radiosurgical fields

Background: To determine the basic physical and dosimetric properties of a new synthetic single-crystal diamond detector and its application for relative small field dosimetry. Materials and Methods: The pre-irradiation dose required to stabilize detector response, dose rate dependence, photon and electron energy dependence, temperature dependence and angular dependence of MicroDiamond detector...

متن کامل

Dose distributions at extreme irradiation depths of gamma knife radiosurgery: EGS4 Monte Carlo calculations.

The accuracy of the dose planning system (Leksell GammaPlan), used in Gamma Knife (type B) radiosurgery at extreme irradiation depths, was verified using the Monte Carlo technique. EGS4 Monte Carlo calculations were employed to calculate the dose distribution along the x, y and z axes for an irradiation relatively shallow in a spherical bony cavity water phantom. Two different sizes of the coll...

متن کامل

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...

متن کامل

Long term effects of Gamma knife Radiosurgery for treatment of cerebral arteriovenous malformations

 Abstract Background: The Gamma Knife Radiosurgery (GKR) is an established management option for Cerebral Ar-teriovenous Malformations (AVMS). Therapeutic benefits of radiosurgery for arteriovenous malformations are complete obliteration of nidus with minimal neurological deficit. Methods: Radiosurgery was performed between February 2003 and April 2010 at Kamraniye day clinic, Teh-ran, Iran, us...

متن کامل

Effects of bone- and air-tissue inhomogeneities on the dose distributions of the Leksell Gamma Knife calculated with PENELOPE.

Monte Carlo simulation with PENELOPE (version 2003) is applied to calculate Leksell Gamma Knife dose distributions for heterogeneous phantoms. The usual spherical water phantom is modified with a spherical bone shell simulating the skull and an air-filled cube simulating the frontal or maxillary sinuses. Different simulations of the 201 source configuration of the Gamma Knife have been carried ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physics in medicine and biology

دوره 43 6  شماره 

صفحات  -

تاریخ انتشار 1998