Gain and maximum output of two electromagnetic middle ear implants: are real ear measurements helpful?

نویسندگان

  • Ad Snik
  • John Noten
  • Cor Cremers
چکیده

We compared the output of two electronic middle ear implants: the Otologics MET device and the Vibrant Soundbridge device. Both devices were programmed in the linear amplification mode. Aided minus unaided sound pressure levels recorded in the ear canal (objective gain) were compared to unaided minus aided soundfield thresholds (functional gain) in 13 patients with severe sensorineural hearing loss. In addition, input/output characteristics were studied with the help of ear canal measurements. Objective gain was consistently lower than functional gain, with wide variation between patients and frequencies. Using input/output data measured in the ear canal in combination with functional gain data, the mean maximum output of the two devices was estimated, expressed in dB SPL. In comparison to NAL-R target values, (functional) gain was adequate; however, the maximum output was low, especially for the Vibrant Soundbridge device.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Testing a method for quantifying the output of implantable middle ear hearing devices.

This report describes tests of a standard practice for quantifying the performance of implantable middle ear hearing devices (also known as implantable hearing aids). The standard and these tests were initiated by the Food and Drug Administration of the United States Government. The tests involved measurements on two hearing devices, one commercially available and the other home built, that wer...

متن کامل

Real ear unaided gain and its relation with the equivalent volume of the external and middle ear

INTRODUCTION  Old age is associated with changes in the characteristics of the middle ear transmission system and in external ear resonance, and these carry implications for the hearing aid (HA) verification process for which targets and measures of the real ear insertion gain (REIG) are used. AIM  To compare the real ear unaided gain (REUG) and the equivalent volumes of the external ear (Veq...

متن کامل

Long-Term Patient Satisfaction With Different Middle Ear Hearing Implants in Sensorineural Hearing Loss

Introduction: Implantable hearing devices are a viable option for patients with moderate-to-severe sensorineural hearing loss who cannot benefit from the conventional hearing aids. In this study, we focus on the patients_ satisfaction with 3 different middle ear implants, the Vibrant SoundBridge (VSB), the semiimplantable Otologics MET implant, and the fully implantable Carina implant. Material...

متن کامل

Simultaneous measurement of middle-ear input impedance and forward/reverse transmission in cat.

Reported here is a technique for measuring forward and reverse middle-ear transmission that exploits distortion-product otoacoustic emissions (DPOAEs) to drive the middle ear "in reverse" without opening the inner ear. The technique allows measurement of DPOAEs, middle-ear input impedance, and forward and reverse middle-ear transfer functions in the same animal. Intermodulation distortion in th...

متن کامل

Middle ear implants: functional gain in mixed hearing loss.

UNLABELLED Osseous atresia and chronic otitis media are diseases benefit with middle ear implants. Surgery for atresia is technically complicated, has significant number of complications and functional results are often poor. The osseointegrated hearing aids are an alternative. They provide a very good functional gain, but have many problems with the skin and osseointegration. In chronic otitis...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of the American Academy of Audiology

دوره 15 3  شماره 

صفحات  -

تاریخ انتشار 2004