Use of a radioactive check device for redundancy check of ionization chambers

نویسندگان

  • N. P. S. Sidhu
  • Alkis Rouvas
  • Patrick Cadman
چکیده

The use of a radioactive check device containing Strontium 90 was investigated to carry out the redundancy checks of Farmer-type ionization chambers. It was not possible to meet the recommended tolerance limits of the redundancy checks without taking into account the angular response of the ionization chambers. The ionization chambers exhibited a maximum variation of 1% in the angular response in this check device. After accounting for the angular response the maximum variation of the short-term repeatability was 0.14% with a standard deviation of +/-0.05%. The long-term constancy checked over a period of nine months was less than +/-0.6% for measurements, taking into account the angular response of the ionization chambers. No significant effect of the background radiation levels was observed on the measurements.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Some steps towards establishing a tertiary standard dosimetry laboratory at a radiotherapy department

Background: In order to deliver the precise dose to the target in radiotherapy, absorbed dose to water at the reference point should be assessed. When the calibration procedure is performed for a reference dosimeter in the 60Co beam of a Secondary Standard Dosimetry Laboratory (SSDL), the total uncertainty in absorbed dose to water (Dw) is estimated to be approximately 1.5%. This study attempts...

متن کامل

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Background: Absolute dosimetry of external beam radiotherapy is carried out by the use of ionization chambers. These chambers must be calibrated at a standard dosimetry laboratory before any use in clinical dosimetry. The secondary standard dosimetry laboratory of Iran (SSDL) has the duty of calibrating the ionization chambers used in radiotherapy centers in Iran. Materials and Methods...

متن کامل

Measurements of output factors using different ionization chambers and build up caps

Background: The aim of this work was to study the output factors for Linac using different ionization chambers and build up caps. Materials and Methods: Output factors were measured for open square fields (3 × 3 cm to 40 × 40 cm) defined by collimator jaws for 6 and 15 MV photon beams from a Varian Clinac 2100C accelerator were measured. The measurements in air were performed using Com...

متن کامل

Cross-Layer Iterative Decoding of Irregular LDPC using Cyclic Redundancy Check Codes

This paper presents a cross layer iterative decoder for irregular Low-Density Parity-Check (LDPC) coded system by using Cyclic Redundancy Check (CRC) codes. The key idea of the decoder is to use the correctly decoded frame to help correcting the left erroneous ones. To accomplish this, the decoder digs the useful information between layers by using the cross layer design method and an iterative...

متن کامل

Design and fabrication of cylindrical ionization chamber for application in radiation protection

According to International Atomic Energy Agency (IAEA) recommendations, the calibration of ionization chambers used for radiotherapy dosimetry follows the substitution method which demands the use of a reference ionization chamber. This work introduces the Design and fabrication of cylindrical ionization chamber for dosimetry to be used by standard dosimetry laboratories. The result of the qual...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 1  شماره 

صفحات  -

تاریخ انتشار 2000