Value of ventricular electrogram recordings in the diagnosis of arrhythmias precipitating electrical device shock therapy.
نویسندگان
چکیده
An antitachycardia pacemaker-cardioverter-defibrillator that is capable of storing ventricular electrograms before and after delivery of device shock therapy was implanted in 16 patients. Three of the patients experienced out-of-hospital device shock therapy preceded by minimal symptoms. Although limitations of electrogram analysis exist and are discussed, careful analysis and registration of electrograms during all supraventricular and ventricular rhythms observed during in-hospital testing served as an important reference for subsequent arrhythmia diagnosis. By analyzing the electrogram rate and RR interval stability and configuration, a definitive diagnosis was established in all three patients (atrial fibrillation, polymorphic ventricular tachycardia and rate-sensing lead disruption, respectively). Thus, the ability to store ventricular electrograms before shock therapy represents a major advance in the management of patients who receive an electrical device to treat ventricular tachyarrhythmia.
منابع مشابه
Implantable cardioverter-defibrillator therapy in the absence of significant symptoms. Rhythm diagnosis and management aided by stored electrogram analysis.
BACKGROUND This report describes the value of stored ventricular electrogram analysis in the diagnosis and management of patients experiencing minimal or no symptoms before implantable cardioverter-defibrillator (ICD) therapy. METHODS AND RESULTS The study population included 48 patients who received the Cadence Tiered Therapy Defibrillator System, an investigational third-generation ICD with...
متن کاملRhythm Diagnosis and Management Aided by Stored Electrogram Analysis
Background. This report describes the value of stored ventricular electrogram analysis in the diagnosis and management of patients experiencing minimal or no symptoms before implantable cardioverterdefibrillator (ICD) therapy. Methods and Results. The study population included 48 patients who received the Cadence Tiered Therapy Defibrillator System, an investigational third-generation ICD with ...
متن کاملImplantable cardioverter-defibrillator in a patient with dextrocardia situs inversus
Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...
متن کاملOccasional dropped ventricular pacing in a patient with no underlying rhythm and an Advisa® dual-chamber pacemaker.
An 86-year-old white male with complete heart block and no underlying ventricular escape rhythm was implanted with an Advisa MRI dual-chamber pacemaker. Telemetry recordings revealed occasional, isolated P waves without pacing spikes or paced QRS complexes. Decreasing the right ventricle (RV) channel sensitivity to a value of 1.2 mV or higher eliminated completely this phenomenon. The observanc...
متن کاملVentricular tachycardia during follow-up in patients resuscitated from ventricular fibrillation: experience from stored electrograms of implantable cardioverter-defibrillators.
OBJECTIVE The purpose of this study was to use the electrogram storage capabilities of the implantable cardioverter-defibrillator (ICD) to categorize any arrhythmic event during follow-up in a group of patients who had survived an episode of ventricular fibrillation (VF) and to possibly identify clinical predictors of future arrhythmic events. BACKGROUND Little is known about the electrophysi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of the American College of Cardiology
دوره 17 4 شماره
صفحات -