Objective analysis of toric intraocular lens rotation and centration.
نویسندگان
چکیده
PURPOSE To assess the repeatability of an objective image-analysis technique to determine intraocular lens (IOL) rotation and centration. SETTING Six European ophthalmology clinics. METHODS Eyes with an Akreos AO aspheric IOL with orientation marks were imaged. A masked observer rated image quality. The axis of rotation was determined from a line bisecting the IOL orientation marks. This was normalized for rotation of the eye between visits using the axis bisecting 2 consistent conjunctival vessels or iris features. The center of ovals overlaid to circumscribe the IOL optic edge and the pupil or limbus were compared to determine IOL centration. Intrasession repeatability was assessed in a subgroup of eyes and the variability of repeated analysis examined. RESULTS The SD of intrasession repeatability was +/-0.79 degrees for IOL rotational stability, +/-0.10 mm for horizontal centration, and +/-0.10 mm for vertical centration. The SD of repeated-analysis variability of the same image was +/-0.70 degrees, +/-0.20 mm, and +/-0.31 mm, respectively. The mean eye rotation (absolute) between visits was 2.23 degrees +/- 1.84 (SD) (10% >5 degrees rotation) using 1 set of consistent conjunctival vessels or iris features and 2.03 +/- 1.66 degrees (7%>5 degrees rotation) using the mean of 2 sets (P = .13). Poorer image quality resulted in larger apparent absolute IOL rotation (r = -0.45, P<.001). CONCLUSIONS Objective analysis of digital retroillumination images allowed sensitive assessment of IOL rotation and centration stability. Eye rotation between images can lead to significant errors if not taken into account. Image quality is important to analysis accuracy.
منابع مشابه
Optical quality of toric intraocular lens implantation in cataract surgery.
AIM To analyze the optical quality after implantation of toric intraocular lens with optical quality analysis system. METHODS Fifty-two eyes of forty-four patients with regular corneal astigmatism of at least 1.00 D underwent implantation of AcrySof toric intraocular lens, including T3 group 19 eyes, T4 group 18 eyes, T5 group 10 eyes, T6 group 5 eyes. Main outcomes evaluated at 3mo of follow...
متن کاملResults of the AcrySof Toric intraocular lenses implantation
BACKGROUND The expectations of post-removal cataract surgery patients are extremely high, and best vision acuity is expected. The best refractive results are influenced by two factors - cataract surgical removal and the corneal astigmatism correction. Currently, the two most often applied corneal astigmatism removal methods are laser surgery and toric intraocular lens implantation, with the lat...
متن کاملRotational stability of the AcrySof SA60TT toric intraocular lenses: A cohort study
BACKGROUND To evaluate the rotational stability of the three types of AcrySof SA60TT toric intraocular lenses (Alcon, Switzerland) in cataract surgery after the first postoperative week. METHODS A retrospective study of 44 eyes in 33 patients. All patients underwent similar uncomplicated phacoemulsification cataract surgery. Seven eyes with corneal astigmatism of less than 1.5 D were implante...
متن کاملAcrylic toric intraocular lens implantation: a single center experience concerning clinical outcomes and postoperative rotation
PURPOSE To present clinical results of toric intraocular lens (IOL) implantation for preexisting astigmatism correction and determine the time of any postoperative rotation. PATIENTS AND METHODS Twenty-nine eyes of 19 patients underwent uncomplicated phacoemulsification and were implanted with an Acrysof ((c)) toric IOL. Uncorrected visual acuity, residual astigmatism, and postoperative rotat...
متن کاملEvaluation of rotation and visual outcomes after implantation of monofocal and multifocal toric intraocular lenses.
PURPOSE To evaluate rotational stability and its influence on postoperative visual acuity of different monofocal and multifocal toric intraocular lenses (IOLs). METHODS A prospective interventional study was designed. Ninety-one patients with a mean age of 71.65 ± 11.82 years were implanted with toric IOLs after phacoemulsification. Three monofocal toric IOLs (the Lentis LT [Oculentis, Berlin...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of cataract and refractive surgery
دوره 36 5 شماره
صفحات -
تاریخ انتشار 2010