Noninvasive pulse CO-oximetry expedites evaluation and management of patients with carbon monoxide poisoning.

نویسنده

  • Neil B Hampson
چکیده

PURPOSES Pulse CO-oximetry (Rad-57; Masimo Corp, Irvine, CA) has been available since 2005. To date, all published clinical studies have focused on clinical reliability and whether the device enhances case finding through screening of various populations. This study examines whether use of pulse CO-oximetry shortens the time to diagnosis and treatment of patients with carbon monoxide (CO) poisoning. BASIC PROCEDURES Data from the joint Undersea and Hyperbaric Medical Society/Centers for Disease Control and Prevention CO poisoning surveillance system from August 2008 to July 2011 were analyzed. Of 1711 cases of CO poisoning treated with hyperbaric oxygen in the United States and reported through the system, 1606 had their initial carboxyhemoglobin (COHb) level measured by laboratory CO-oximetry and 105 by pulse CO-oximetry. Patients were selected from the laboratory CO-oximetry group to match each of the 105 patients evaluated by pulse CO-oximetry in 5 characteristics--age, sex, race/ethnicity, intent of poisoning, and occurrence of loss of consciousness. Measures of timeliness in measurement and management were compared between the 2 groups. MAIN FINDINGS Patients with initial COHb measurement by pulse CO-oximetry had significantly shorter time to measurement of COHb, higher average levels of COHb, and shorter time from the end of CO exposure to the initiation of hyperbaric oxygen treatment. On average, patients evaluated by pulse CO-oximetry reached the hyperbaric chamber 1 hour faster than did patients evaluated by laboratory CO-oximetry (P < .01). PRINCIPLE CONCLUSIONS Pulse CO-oximetry is associated with more rapid diagnosis and initiation of hyperbaric oxygen therapy in CO-poisoned patients compared with laboratory CO-oximetry. The impact on clinical outcome remains to be determined.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of noninvasive multiwave pulse oximetry compared with carboxyhemoglobin from blood gas analysis in unselected emergency department patients.

STUDY OBJECTIVE Accurate and timely diagnosis of carbon monoxide (CO) poisoning is difficult because of nonspecific symptoms. Multiwave pulse oximetry might facilitate the screening for occult poisoning by noninvasive measurement of carboxyhemoglobin (COHb), but its reliability is still unknown. We assess bias and precision of COHb oximetry compared with the criterion standard blood gas analysi...

متن کامل

Emergency department management of suspected carbon monoxide poisoning: role of pulse CO-oximetry.

BACKGROUND The RAD-57 pulse CO-oximeter is a lightweight device allowing noninvasive measurement of blood carboxyhemoglobin (S(pCO)). We assessed the diagnostic value of pulse CO-oximetry, comparing S(pCO) values from the RAD-57 to standard laboratory blood carboxyhemoglobin (COHb) measurement in emergency department patients with suspected carbon monoxide (CO) poisoning. METHODS This was a p...

متن کامل

Reliability of new pulse CO-oximeter in victims of carbon monoxide poisoning.

STUDY OBJECTIVE The purpose of this study was to evaluate the reliability of noninvasive real-time measurement of carboxyhemoglobin (COHb) using a pulse CO-oximeter in victims of carbon monoxide poisoning (COP). METHODS During the 7-month study period, pulse CO-oximetry was measured on patients admitted to the emergency department (ED) for suspected COP. Each patient included in the study und...

متن کامل

Management of carbon monoxide poisoning using oxygen therapy.

The management of carbon monoxide poisoning requires an accurate assessment of the extent of blood oxygenation. Measuring the fractional oxyhaemoglobin content by using co-oximetry gives a true picture of the oxygen-carrying capacity of blood in the presence of carboxyhaemoglobin. The use of readings from pulse oximetry or a standard blood gas analyser is insufficient and can be misleading. We ...

متن کامل

Victim of carbon monoxide poisoning identified by carbon monoxide oximetry.

BACKGROUND Carbon monoxide (CO) is known as a leading cause of unintentional poisoning death in many countries. Diagnosis is usually made by measuring carboxyhemoglobin (COHb), but due to the non-specific symptoms, the crucial step is considering CO poisoning. A possible solution might be screening emergency department (ED) patients. A cutaneous sensor that measures oxygen and CO saturation sim...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The American journal of emergency medicine

دوره 30 9  شماره 

صفحات  -

تاریخ انتشار 2012