Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy
نویسندگان
چکیده
BACKGROUND Volumetric-modulated Dynamic WaveArc therapy (VMDWAT) is a non-coplanar continuous volumetric modulated radiation therapy (VMAT) delivery technique. Here, we monitored mechanical errors and their impact on dose distributions in VMDWAT using logfiles throughout the course of treatment. METHODS Fifteen patients were enrolled (2 skull base tumor patients and 13 prostate cancer patients). VMDWAT plans were created for the enrolled patients. The prescribed dose for the skull base tumor was set as 54 Gy at 1.8 Gy per fraction, and that for the prostate cancer was set as 72 to 78 Gy at 2 Gy per fraction. We acquired logfiles to monitor mechanical errors and their impact on dose distribution in each fraction. The root mean square error (RMSE) in the multi-leaf collimator (MLC), gantry angle, O-ring angle and monitor unit (MU) were calculated using logfiles throughout the course of VMDWAT for each patient. The dosimetric impact of mechanical errors throughout the course of VMDWAT was verified using a logfile-based dose reconstruction method. Dosimetric errors between the reconstructed plans and the original plans were assessed. RESULTS A total of 517 datasets, including 55 datasets for the 2 skull base tumor patients and 462 datasets for the 13 prostate cancer patients, were acquired. The RMSE values were less than 0.1 mm, 0.2°, 0.1°, and 0.4 MU for MLC position, gantry angle, O-ring angle, and MU, respectively. For the skull base tumors, the absolute mean dosimetric errors and two standard deviations throughout the course of treatment were less than 1.4% and 1.1%, respectively. For prostate cancer, these absolute values were less than 0.3% and 0.5%, respectively. The largest dosimetric error of 2.5% was observed in a skull base tumor patient. The resultant dosimetric error in the accumulated daily delivered dose distribution, in the patient with the largest error, was up to 1.6% for all dose-volumetric parameters relative to the planned dose distribution. CONCLUSIONS MLC position, gantry rotation, O-ring rotation and MU were highly accurate and stable throughout the course of treatment. The daily dosimetric errors due to mechanical errors were small. VMDWAT provided high delivery accuracy and stability throughout the course of treatment. TRIAL REGISTRATION UMIN000023870 . Registered: 1 October 2016.
منابع مشابه
Quantification and modelling of the dosimetric impact of the treatment couch in volumetric modulated arc therapy (VMAT)
Background: As the volumetric modulated arc therapy (VMAT) becoming a main role of treatment ways, the effect of couch top becomes more significant. It is imperative to re-evaluate the couches that previously may have been considered of no importance during early treatment techniques. The impact of couch top on radiation delivery was explored and the couch model was tested with the aim of reduc...
متن کاملDosimetric Effects of Rotational Setup Error in Volumetric Modulated Arc Radiotherapy on Brain Tumor Patients
Introduction: This study examined the dosimetric effects based on the rotational setup error to correct patient setup errors occur during volumetric modulated arc radiotherapy (VMAT) for brain tumor patients. Material and Methods: This study included 1129 cases of cone beam computed tomography (CBCT) images obtained from 46 brain tumor patients, who experienced VMAT and used the 6DoF (degree o...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملA dosimetric comparison of volumetric modulated arc therapy (VMAT) and non-coplanar intensity modulated radiotherapy (IMRT) for nasal cavity and paranasal sinus cancer
BACKGROUND To compare dosimetric parameters of volumetric modulated arc therapy (VMAT) and non-coplanar intensity modulated radiotherapy (IMRT) for nasal cavity and paranasal sinus cancer with regard to the coverage of planning target volume (PTV) and the sparing of organs at risk (OAR). METHODS Ten patients with nasal cavity or paranasal sinus cancer were re-planned by VMAT (two-arc) plan an...
متن کاملComparison of Volumetric Modulated Arc Therapy and Three-Dimensional Conformal Radiotherapy in Postoperative High-Grade Glioma: A Dosimetric Comparison
Introduction: We aimed to dosimetrically compare three-dimensional conformal radiotherapy (3D-CRT) and volumetric modulated arc therapy (VMAT) in terms of planning target volume (PTV) coverage, organ at risk (OAR) sparing, and conformity index (CI). Material and Methods: Planning data of 26 high grade glioma (HGG) patients were used. Prescrib...
متن کامل